We present the realization of a prototype of a compact, switchable electron source operating at cryogenic temperatures, demonstrated at energies ranging from 100 eV to 300 eV, and conceptually extendable to an arbitrary energy range. The electrons are produced via photoelectric emission, induced by a LED illuminating two 400-nm-thick commercial aluminum layers inside a cryostat, and are subsequently accelerated by a voltage of up to 300 V. Detection is carried out with an array of transition-edge sensor (TES) microcalorimeters designed for X-ray detectors, where we successfully observe signals consistent with electrons produced at the source at rates of 1 Hz and with an efficiency of .

Bennett, D., Borghesi, M., Campana, P., Carobene, R., Cattaneo, A., De Gerone, M., et al. (2026). Demonstration of a cryogenic, switchable electron source for low-temperature detector calibration. THE EUROPEAN PHYSICAL JOURNAL PLUS, 141(10) [10.1140/epjp/s13360-026-08330-z].

Demonstration of a cryogenic, switchable electron source for low-temperature detector calibration

Borghesi, M.;Campana, P.;Carobene, R.;Cattaneo, A.;Faverzani, M.;Ferri, E.;Gamba, S.;Giachero, A.;Gobbo, M.;Labranca, D.;Nucciotti, A.;
2026

Abstract

We present the realization of a prototype of a compact, switchable electron source operating at cryogenic temperatures, demonstrated at energies ranging from 100 eV to 300 eV, and conceptually extendable to an arbitrary energy range. The electrons are produced via photoelectric emission, induced by a LED illuminating two 400-nm-thick commercial aluminum layers inside a cryostat, and are subsequently accelerated by a voltage of up to 300 V. Detection is carried out with an array of transition-edge sensor (TES) microcalorimeters designed for X-ray detectors, where we successfully observe signals consistent with electrons produced at the source at rates of 1 Hz and with an efficiency of .
Articolo in rivista - Articolo scientifico
low temperature detectors, low-energy electrons
English
14-ott-2026
2026
141
10
1176
open
Bennett, D., Borghesi, M., Campana, P., Carobene, R., Cattaneo, A., De Gerone, M., et al. (2026). Demonstration of a cryogenic, switchable electron source for low-temperature detector calibration. THE EUROPEAN PHYSICAL JOURNAL PLUS, 141(10) [10.1140/epjp/s13360-026-08330-z].
File in questo prodotto:
File Dimensione Formato  
Bennett et al-2026-Eur. Phys. J. Plus-VoR.pdf

accesso aperto

Tipologia di allegato: Publisher’s Version (Version of Record, VoR)
Licenza: Creative Commons
Dimensione 1.55 MB
Formato Adobe PDF
1.55 MB Adobe PDF Visualizza/Apri