The solid state galvanic chain Pt, Ag|(Ag2SO4)25-(Li2SO4)75|SO2, O2(air), Pt is used in order to study the response time of an SO2 sensor as a function of the configuration of the measuring electrode as well as the thickness of the electrolyte. A discussion concerning the influence of both parameters on the response time is presented and a proposal to improve the performance of the sensor is suggested
Mari, C., Beghi, M., Pizzini, S., Faltemier, J. (1990). Electrochemical solid state sensor for SO2 determination in air. SENSORS AND ACTUATORS. B, CHEMICAL, 2(1), 51-55 [10.1016/0925-4005(90)80008-N].
Electrochemical solid state sensor for SO2 determination in air
MARI, CLAUDIO MARIA;Pizzini, S;
1990
Abstract
The solid state galvanic chain Pt, Ag|(Ag2SO4)25-(Li2SO4)75|SO2, O2(air), Pt is used in order to study the response time of an SO2 sensor as a function of the configuration of the measuring electrode as well as the thickness of the electrolyte. A discussion concerning the influence of both parameters on the response time is presented and a proposal to improve the performance of the sensor is suggestedFile in questo prodotto: