A feasibility study for the use of a Ce3+-doped optical fibre as a radio luminescent dosimeter in radiotherapy was carried out. The prototype showed a satisfactory reproducibility (1.7%) and a good linearity over a clinically relevant dose range (from few cGy to approximately 10 Gy). Moreover, the device enabled a reliable evaluation of the absorbed dose, independently of the dose rate and of the orientation of the incoming radiation. A slight energy dependence of the response was observed when the system was irradiated with electron beams of different energies. The results indicate that the employment of this new system might be very promising both for quality control measurements and "in vivo" dosimetry. Some improvements are nevertheless still required in order to allow a real-time evaluation of the Cherenkov radiation produced by the irradiated portion of the passive optical fibre, which influences some of the dosimetric properties of the system. (c) 2006 Elsevier B.V. All rights reserved.
Mones, E., Veronese, I., Moretti, F., Fasoli, M., Loi, G., Negri, E., et al. (2006). Feasibility study for the use of Ce3+-doped optical fibres in radiotherapy. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH. SECTION A, ACCELERATORS, SPECTROMETERS, DETECTORS AND ASSOCIATED EQUIPMENT, 562(1), 449-455 [10.1016/j.nima.2006.02.003].
Feasibility study for the use of Ce3+-doped optical fibres in radiotherapy
MORETTI, FEDERICO;FASOLI, MAURO;CHIODINI, NORBERTO;VEDDA, ANNA GRAZIELLA
2006
Abstract
A feasibility study for the use of a Ce3+-doped optical fibre as a radio luminescent dosimeter in radiotherapy was carried out. The prototype showed a satisfactory reproducibility (1.7%) and a good linearity over a clinically relevant dose range (from few cGy to approximately 10 Gy). Moreover, the device enabled a reliable evaluation of the absorbed dose, independently of the dose rate and of the orientation of the incoming radiation. A slight energy dependence of the response was observed when the system was irradiated with electron beams of different energies. The results indicate that the employment of this new system might be very promising both for quality control measurements and "in vivo" dosimetry. Some improvements are nevertheless still required in order to allow a real-time evaluation of the Cherenkov radiation produced by the irradiated portion of the passive optical fibre, which influences some of the dosimetric properties of the system. (c) 2006 Elsevier B.V. All rights reserved.I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.