The Physics Department of the Technical University of Munich operates a shallow underground detector laboratory in Garching, Germany. It provides ∼160m2 of laboratory space which is shielded from cosmic radiation by ∼6m of gravel and soil, corresponding to a shielding of ∼15m.w.e. The laboratory also houses a cleanroom equipped with work- and wetbenches, a chemical fumehood as well as a spin-coater and a mask-aligner for photolithographic processing of semiconductor detectors. Furthermore, the shallow underground laboratory runs two high-purity germanium detector screening stations, a liquid argon cryostat and a 3He–4He dilution refrigerator with a base temperature of ≤12-14mK. The infrastructure provided by the shallow laboratory is particularly relevant for the characterization of CaWO 4 target crystals for the CRESST-III experiment, detector fabrication and assembly for rare event searches. Future applications of the laboratory include detector development in the framework of coherent neutrino nucleus scattering experiments (ν-cleus) and studying its potential as a site to search for MeV-scale dark matter with gram-scale cryogenic detectors.

Langenkamper, A., Defay, X., Ferreiro Iachellini, N., Kinast, A., Lanfranchi, J., Lindner, E., et al. (2018). A Cryogenic Detector Characterization Facility in the Shallow Underground Laboratory at the Technical University of Munich. JOURNAL OF LOW TEMPERATURE PHYSICS, 193(5-6), 860-866 [10.1007/s10909-018-1899-y].

A Cryogenic Detector Characterization Facility in the Shallow Underground Laboratory at the Technical University of Munich

Ferreiro Iachellini N.;
2018

Abstract

The Physics Department of the Technical University of Munich operates a shallow underground detector laboratory in Garching, Germany. It provides ∼160m2 of laboratory space which is shielded from cosmic radiation by ∼6m of gravel and soil, corresponding to a shielding of ∼15m.w.e. The laboratory also houses a cleanroom equipped with work- and wetbenches, a chemical fumehood as well as a spin-coater and a mask-aligner for photolithographic processing of semiconductor detectors. Furthermore, the shallow underground laboratory runs two high-purity germanium detector screening stations, a liquid argon cryostat and a 3He–4He dilution refrigerator with a base temperature of ≤12-14mK. The infrastructure provided by the shallow laboratory is particularly relevant for the characterization of CaWO 4 target crystals for the CRESST-III experiment, detector fabrication and assembly for rare event searches. Future applications of the laboratory include detector development in the framework of coherent neutrino nucleus scattering experiments (ν-cleus) and studying its potential as a site to search for MeV-scale dark matter with gram-scale cryogenic detectors.
Articolo in rivista - Articolo scientifico
Cryogenic detectors; Detector characterization; Low-background techniques; Shallow underground laboratory;
English
2018
193
5-6
860
866
none
Langenkamper, A., Defay, X., Ferreiro Iachellini, N., Kinast, A., Lanfranchi, J., Lindner, E., et al. (2018). A Cryogenic Detector Characterization Facility in the Shallow Underground Laboratory at the Technical University of Munich. JOURNAL OF LOW TEMPERATURE PHYSICS, 193(5-6), 860-866 [10.1007/s10909-018-1899-y].
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/10281/628779
Citazioni
  • Scopus 7
  • ???jsp.display-item.citation.isi??? 7
Social impact