Background: This investigation used image data generated by a physical phantom over a wide range of count statistics to evaluate the effectiveness of several of the newer commercially available SPECT reconstruction iterative algorithms (IRR) in improving perfusion defect contrast and spatial resolution, while controlling image noise. Methods: A cardiac phantom was imaged using four different gamma cameras over a wide range of counts statistics (from 6 to 0.8 Mcounts). Images were reconstructed with FBP, OSEM, and the IRR available on site. IRR were applied without corrections (IRR NC), with attenuation correction (IRR AC), scatter correction (IRR SC), and attenuation + scatter corrections (IRR SCAC). Four image performance indices related to spatial resolution, contrast, and image noise were analyzed. Results: IRR NC always determined significant improvements in all indices in comparison to FBP or OSEM. Improvements were emphasized with IRR SC and IRR SCAC. Count reduction from 6 to 1.5 Mcounts did not impair the performances of any of the considered indices. Conclusions: This is the first study comparing the relative performance of different, commercially available, IRR software, over a wide range of count statistics; the additional effect of scatter and attenuation corrections, alone or in combination, was also evaluated. Our results confirm that IRR algorithms produce substantial benefits with respect to conventional FBP or OSEM reconstruction methods, as assessed through different figures of merit, in particular when SC and/or SCAC are also included. © 2013 American Society of Nuclear Cardiology.

Zoccarato, O., Scabbio, C., DE PONTI, E., Matheoud, R., Leva, L., Morzenti, S., et al. (2014). Comparative analysis of iterative reconstruction algorithms with resolution recovery for cardiac SPECT studies. A multi-center phantom study. JOURNAL OF NUCLEAR CARDIOLOGY, 21(1), 135-148 [10.1007/s12350-013-9821-0].

Comparative analysis of iterative reconstruction algorithms with resolution recovery for cardiac SPECT studies. A multi-center phantom study

DE PONTI, ELENA;MORZENTI, SABRINA;CRIVELLARO, CINZIA;
2014

Abstract

Background: This investigation used image data generated by a physical phantom over a wide range of count statistics to evaluate the effectiveness of several of the newer commercially available SPECT reconstruction iterative algorithms (IRR) in improving perfusion defect contrast and spatial resolution, while controlling image noise. Methods: A cardiac phantom was imaged using four different gamma cameras over a wide range of counts statistics (from 6 to 0.8 Mcounts). Images were reconstructed with FBP, OSEM, and the IRR available on site. IRR were applied without corrections (IRR NC), with attenuation correction (IRR AC), scatter correction (IRR SC), and attenuation + scatter corrections (IRR SCAC). Four image performance indices related to spatial resolution, contrast, and image noise were analyzed. Results: IRR NC always determined significant improvements in all indices in comparison to FBP or OSEM. Improvements were emphasized with IRR SC and IRR SCAC. Count reduction from 6 to 1.5 Mcounts did not impair the performances of any of the considered indices. Conclusions: This is the first study comparing the relative performance of different, commercially available, IRR software, over a wide range of count statistics; the additional effect of scatter and attenuation corrections, alone or in combination, was also evaluated. Our results confirm that IRR algorithms produce substantial benefits with respect to conventional FBP or OSEM reconstruction methods, as assessed through different figures of merit, in particular when SC and/or SCAC are also included. © 2013 American Society of Nuclear Cardiology.
Articolo in rivista - Articolo scientifico
Iterative reconstruction; myocardial perfusion imaging; reconstruction algorithms; resolution recovery; SPECT; Cardiology and Cardiovascular Medicine; Radiology, Nuclear Medicine and Imaging
English
2014
21
1
135
148
reserved
Zoccarato, O., Scabbio, C., DE PONTI, E., Matheoud, R., Leva, L., Morzenti, S., et al. (2014). Comparative analysis of iterative reconstruction algorithms with resolution recovery for cardiac SPECT studies. A multi-center phantom study. JOURNAL OF NUCLEAR CARDIOLOGY, 21(1), 135-148 [10.1007/s12350-013-9821-0].
File in questo prodotto:
File Dimensione Formato  
zoccarato.pdf

Solo gestori archivio

Dimensione 788.21 kB
Formato Adobe PDF
788.21 kB Adobe PDF   Visualizza/Apri   Richiedi una copia

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/10281/56577
Citazioni
  • Scopus 34
  • ???jsp.display-item.citation.isi??? 32
Social impact