A new method for the calibration of multi-axis, linear and redundant force sensors is presented. This new procedure, named device hyperplane characterization method, is inspired by the shape from motion method for it reduces the burden represented by the need for a huge number of force measurements, typical using least-squares methods, in order to reject errors during the calibration procedure. The proposed technique is an application of the rank theorem and achieves good noise rejection without much time consumption focusing on sensor output measurements, and reducing the effect of disturbances operating the projection of raw output data on the hyperplane to which measurements are ideally compelled to belong in the case of redundant sensors.

Oddo, C., Valdastri, P., Beccai, L., Roccella, S., Carrozza, M., Dario, P. (2007). Investigation on calibration methods for multi-axis, linear and redundant force sensors. MEASUREMENT SCIENCE & TECHNOLOGY, 18(3), 623-631 [10.1088/0957-0233/18/3/011].

Investigation on calibration methods for multi-axis, linear and redundant force sensors

Carrozza M. C.;
2007

Abstract

A new method for the calibration of multi-axis, linear and redundant force sensors is presented. This new procedure, named device hyperplane characterization method, is inspired by the shape from motion method for it reduces the burden represented by the need for a huge number of force measurements, typical using least-squares methods, in order to reject errors during the calibration procedure. The proposed technique is an application of the rank theorem and achieves good noise rejection without much time consumption focusing on sensor output measurements, and reducing the effect of disturbances operating the projection of raw output data on the hyperplane to which measurements are ideally compelled to belong in the case of redundant sensors.
Articolo in rivista - Articolo scientifico
Calibration; Hyperplane; Linear; Redundant; Triaxial force sensor;
English
2007
18
3
623
631
011
reserved
Oddo, C., Valdastri, P., Beccai, L., Roccella, S., Carrozza, M., Dario, P. (2007). Investigation on calibration methods for multi-axis, linear and redundant force sensors. MEASUREMENT SCIENCE & TECHNOLOGY, 18(3), 623-631 [10.1088/0957-0233/18/3/011].
File in questo prodotto:
File Dimensione Formato  
Oddo-2007-Measurement Science and Technology-VoR.pdf

Solo gestori archivio

Tipologia di allegato: Publisher’s Version (Version of Record, VoR)
Licenza: Tutti i diritti riservati
Dimensione 338.69 kB
Formato Adobe PDF
338.69 kB Adobe PDF   Visualizza/Apri   Richiedi una copia

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/10281/559368
Citazioni
  • Scopus 45
  • ???jsp.display-item.citation.isi??? 42
Social impact