OBJECTIVE: The aim of this study was to evaluate the robustness and zero-drift of an intracranial pressure sensor, Neurovent-P (Raumedic AG, Münchberg, Germany), when used in the clinical environment. METHODS: A prospective multicenter trial, conforming to the International Organization for Standardization 14155 Standard, was conducted in 6 European BrainIT centers between July 2005 and December 2006. Ninety-nine catheters were used. The study was observational, followed by a centralized sensor bench test after catheter removal. RESULTS: The mean recorded value before probe insertion was 0.17 ± 1.1 mm Hg. Readings outside the range ±1 mm Hg were recorded in only 3 centers on a total of 15 catheters. Complications were minimal and mainly related to the insertion bolt. The mean recorded pressure value at removal was 0.8 ± 2.2 mm Hg. No relationship was identified between postremoval reading and length of monitoring. The postremoval bench test indicated the probability of a system failure, defined as a drift of more than 3 mm Hg, at a range between 12 and 17%. CONCLUSION: The Neurovent-P catheter performed well in clinical use in terms of robustness. The majority of technical complications were associated with the bolt fixation technology. Adverse events were rare and clinically nonsignificant. Despite the earlier reported excellent bench test zero-drift rates, under the more demanding clinical conditions, zero-drift rate remains a concern with catheter tip strain gauge technology. This performance is similar, and not superior, to other intracranial pressure devices.

Citerio, G., Piper, I., Chambers, I., Galli, D., Enblad, P., Kiening, K., et al. (2008). Multicenter clinical assessment of the raumedic Neurovent-P intracranial pressure sensor: A report by the brainIT group. NEUROSURGERY, 63(6), 1152-1158 [10.1227/01.NEU.0000335148.87042.D7].

Multicenter clinical assessment of the raumedic Neurovent-P intracranial pressure sensor: A report by the brainIT group

Citerio, G
;
2008

Abstract

OBJECTIVE: The aim of this study was to evaluate the robustness and zero-drift of an intracranial pressure sensor, Neurovent-P (Raumedic AG, Münchberg, Germany), when used in the clinical environment. METHODS: A prospective multicenter trial, conforming to the International Organization for Standardization 14155 Standard, was conducted in 6 European BrainIT centers between July 2005 and December 2006. Ninety-nine catheters were used. The study was observational, followed by a centralized sensor bench test after catheter removal. RESULTS: The mean recorded value before probe insertion was 0.17 ± 1.1 mm Hg. Readings outside the range ±1 mm Hg were recorded in only 3 centers on a total of 15 catheters. Complications were minimal and mainly related to the insertion bolt. The mean recorded pressure value at removal was 0.8 ± 2.2 mm Hg. No relationship was identified between postremoval reading and length of monitoring. The postremoval bench test indicated the probability of a system failure, defined as a drift of more than 3 mm Hg, at a range between 12 and 17%. CONCLUSION: The Neurovent-P catheter performed well in clinical use in terms of robustness. The majority of technical complications were associated with the bolt fixation technology. Adverse events were rare and clinically nonsignificant. Despite the earlier reported excellent bench test zero-drift rates, under the more demanding clinical conditions, zero-drift rate remains a concern with catheter tip strain gauge technology. This performance is similar, and not superior, to other intracranial pressure devices.
Articolo in rivista - Articolo scientifico
Intracranial pressure; Multicenter trials; Physiological monitoring; Technology assessment;
English
2008
63
6
1152
1158
reserved
Citerio, G., Piper, I., Chambers, I., Galli, D., Enblad, P., Kiening, K., et al. (2008). Multicenter clinical assessment of the raumedic Neurovent-P intracranial pressure sensor: A report by the brainIT group. NEUROSURGERY, 63(6), 1152-1158 [10.1227/01.NEU.0000335148.87042.D7].
File in questo prodotto:
File Dimensione Formato  
Neurosurgery 2008 Citerio.pdf

Solo gestori archivio

Dimensione 490.24 kB
Formato Adobe PDF
490.24 kB Adobe PDF   Visualizza/Apri   Richiedi una copia

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/10281/54979
Citazioni
  • Scopus 43
  • ???jsp.display-item.citation.isi??? 38
Social impact