Objectives: To compare image quality and radiation dose of CT images reconstructed with model-based iterative reconstruction (MBIR) and hybrid-iterative (HIR) algorithm in oncologic patients. Methods: 125 oncologic patients underwent both contrast-enhanced low- (100 kV), and standard (120 kV) dose CT, were enrolled. Image quality was assessed by using a 4-point Likert scale. CT attenuation values, expressed in Hounsfield unit (HU), were recorded within a regions of interest (ROI) of liver, spleen, paraspinal muscle, aortic lumen, and subcutaneous fat tissue. Image noise, expressed as standard deviation (SD), signal-to-noise ratio (SNR), and contrast-to-noise ratio (CNR) were calculated. Radiation dose were analyzed. Paired Student’s t-test was used to compare all continuous variables. Results: The overall median score assessed as image quality for CT images with the MBIR algorithm was significantly higher in comparison with HIR [4 (range 3–4) vs 3 (3-4), p = 0.017]. CT attenuation values and SD were significantly higher and lower, respectively, in all anatomic districts in images reconstructed with MBIR in comparison with HIR ones (all p < 0.001). SNR and CNR values were higher in CT images reconstructed with MBIR, reaching a significant difference in all districts (all p < 0.001). Radiation dose were significantly lower in the MBIR group compared with the HIR group (p < 0.001). Conclusions: MBIR combined with low-kV setting allows an important dose reduction in whole-body CT imaging, reaching a better image quality both qualitatively and quantitatively. Advances in knowledge: MBIR with low-dose approach allows a reduction of dose exposure, maintaining high image quality, especially in patients which deserve a longlasting follow-up.

Ippolito, D., Maino, C., Pecorelli, A., Salemi, I., Gandola, D., Riva, L., et al. (2021). Application of low-dose CT combined with model-based iterative reconstruction algorithm in oncologic patients during follow-up: dose reduction and image quality. BRITISH JOURNAL OF RADIOLOGY, 94(1124) [10.1259/bjr.20201223].

Application of low-dose CT combined with model-based iterative reconstruction algorithm in oncologic patients during follow-up: dose reduction and image quality

Ippolito, D;Gandola, D;Sironi, S
2021

Abstract

Objectives: To compare image quality and radiation dose of CT images reconstructed with model-based iterative reconstruction (MBIR) and hybrid-iterative (HIR) algorithm in oncologic patients. Methods: 125 oncologic patients underwent both contrast-enhanced low- (100 kV), and standard (120 kV) dose CT, were enrolled. Image quality was assessed by using a 4-point Likert scale. CT attenuation values, expressed in Hounsfield unit (HU), were recorded within a regions of interest (ROI) of liver, spleen, paraspinal muscle, aortic lumen, and subcutaneous fat tissue. Image noise, expressed as standard deviation (SD), signal-to-noise ratio (SNR), and contrast-to-noise ratio (CNR) were calculated. Radiation dose were analyzed. Paired Student’s t-test was used to compare all continuous variables. Results: The overall median score assessed as image quality for CT images with the MBIR algorithm was significantly higher in comparison with HIR [4 (range 3–4) vs 3 (3-4), p = 0.017]. CT attenuation values and SD were significantly higher and lower, respectively, in all anatomic districts in images reconstructed with MBIR in comparison with HIR ones (all p < 0.001). SNR and CNR values were higher in CT images reconstructed with MBIR, reaching a significant difference in all districts (all p < 0.001). Radiation dose were significantly lower in the MBIR group compared with the HIR group (p < 0.001). Conclusions: MBIR combined with low-kV setting allows an important dose reduction in whole-body CT imaging, reaching a better image quality both qualitatively and quantitatively. Advances in knowledge: MBIR with low-dose approach allows a reduction of dose exposure, maintaining high image quality, especially in patients which deserve a longlasting follow-up.
Articolo in rivista - Articolo scientifico
Adult; Aftercare; Aged; Female; Humans; Male; Middle Aged; Neoplasms; Retrospective Studies; Tomography, X-Ray Computed; Algorithms; Image Processing, Computer-Assisted; Radiation Dosage
English
2021
94
1124
20201223
none
Ippolito, D., Maino, C., Pecorelli, A., Salemi, I., Gandola, D., Riva, L., et al. (2021). Application of low-dose CT combined with model-based iterative reconstruction algorithm in oncologic patients during follow-up: dose reduction and image quality. BRITISH JOURNAL OF RADIOLOGY, 94(1124) [10.1259/bjr.20201223].
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/10281/472867
Citazioni
  • Scopus 4
  • ???jsp.display-item.citation.isi??? 4
Social impact