High impedance connecting links and cables are tested at low frequency in terms of their parasitic impedance to ground and to neighboring connecting links. These parameters must be well characterized with detectors operated at low temperature, especially when the very front-end is at room temperature, which results in a long link. This is the case of the LUCIFER experiment, an array of crystals where every event of interest produces two signals, one composed of phonons, the other of photons. The parasitic impedance is usually considered to be the parallel combination of a resistance and a capacitance. We characterized both and found that from the static measurements the capacitance of the cable resulted much larger. On the basis of this result we optimized the measurement set-up and developed a model to account for this behavior
Giachero, A., Gotti, C., Maino, M., Pessina, G. (2013). Characterization of high impedance connecting links for Bolometric detectors. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH. SECTION A, ACCELERATORS, SPECTROMETERS, DETECTORS AND ASSOCIATED EQUIPMENT, 718, 229-230 [10.1016/j.nima.2012.11.140].
Characterization of high impedance connecting links for Bolometric detectors
GIACHERO, ANDREA;GOTTI, CLAUDIO;MAINO, MATTEO;Pessina, G.
2013
Abstract
High impedance connecting links and cables are tested at low frequency in terms of their parasitic impedance to ground and to neighboring connecting links. These parameters must be well characterized with detectors operated at low temperature, especially when the very front-end is at room temperature, which results in a long link. This is the case of the LUCIFER experiment, an array of crystals where every event of interest produces two signals, one composed of phonons, the other of photons. The parasitic impedance is usually considered to be the parallel combination of a resistance and a capacitance. We characterized both and found that from the static measurements the capacitance of the cable resulted much larger. On the basis of this result we optimized the measurement set-up and developed a model to account for this behaviorI documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.