BACKGROUND: We assessed the reliability and feasibility of blood pressure (BP) measurements by means of a new wearable watch-type BP monitor (HeartGuide) in detecting episodes of hypotensive (pre)syncope induced by tilt table test. METHODS AND RESULTS: An intrapatient comparison between systolic BP (SBP) measured by means of the HeartGuide device and noninvasive finger beat-to-beat BP monitoring was undertaken both at baseline in supine position and repeatedly during tilt table test in patients evaluated for reflex syncope. Intrapatient fall of systolic BP from baseline was measured. Eighty-one patients (mean age, 61±19 years; 46 women) were included. Overall, HeartGuide was able to yield BP values at the time of BP nadir in 58 (72%) patients (average HeartGuide SBP 102±18 mm Hg, versus finger SBP 101±19 mm Hg). Compared with baseline, the maximum SBP decrease was on average −28.5±27.8 and −30.3±33.9 mm Hg respectively (Lin’s concordance correlation coefficient=0.78, r=0.79, P=0.001). In the subgroup of 38 patients with tilt table test induced (pre)syncope, the average HeartGuide SBP during symptoms was 97±16 mm Hg, and the finger SBP was 94±18 mm Hg. Compared with baseline, the maximum SBP decrease was on average −35.2±29.3 and −43.3±31.8 mm Hg, respectively (Lin’s concordance correlation coefficient=0.83, r=0.87, P=0.001). CONCLUSIONS: Our data indicate that the HeartGuide BP monitor can detect low BP during presyncope and that its measure of SBP change is consistent with that simultaneously obtained through continuous BP monitoring, despite some intrapatient variability. Thus, this device might be useful in determining the hypotensive nature of spontaneous (pre)syncopal symptoms, a possibility that should be verified by field studies.

Groppelli, A., Rafanelli, M., Testa, G., Agusto, S., Rivasi, G., Ungar, A., et al. (2022). Feasibility of Blood Pressure Measurement With a Wearable (Watch-Type) Monitor During Impending Syncopal Episodes. JOURNAL OF THE AMERICAN HEART ASSOCIATION. CARDIOVASCULAR AND CEREBROVASCULAR DISEASE, 11(16) [10.1161/JAHA.122.026420].

Feasibility of Blood Pressure Measurement With a Wearable (Watch-Type) Monitor During Impending Syncopal Episodes

Soranna D.;Zambon A.;Parati G.
2022

Abstract

BACKGROUND: We assessed the reliability and feasibility of blood pressure (BP) measurements by means of a new wearable watch-type BP monitor (HeartGuide) in detecting episodes of hypotensive (pre)syncope induced by tilt table test. METHODS AND RESULTS: An intrapatient comparison between systolic BP (SBP) measured by means of the HeartGuide device and noninvasive finger beat-to-beat BP monitoring was undertaken both at baseline in supine position and repeatedly during tilt table test in patients evaluated for reflex syncope. Intrapatient fall of systolic BP from baseline was measured. Eighty-one patients (mean age, 61±19 years; 46 women) were included. Overall, HeartGuide was able to yield BP values at the time of BP nadir in 58 (72%) patients (average HeartGuide SBP 102±18 mm Hg, versus finger SBP 101±19 mm Hg). Compared with baseline, the maximum SBP decrease was on average −28.5±27.8 and −30.3±33.9 mm Hg respectively (Lin’s concordance correlation coefficient=0.78, r=0.79, P=0.001). In the subgroup of 38 patients with tilt table test induced (pre)syncope, the average HeartGuide SBP during symptoms was 97±16 mm Hg, and the finger SBP was 94±18 mm Hg. Compared with baseline, the maximum SBP decrease was on average −35.2±29.3 and −43.3±31.8 mm Hg, respectively (Lin’s concordance correlation coefficient=0.83, r=0.87, P=0.001). CONCLUSIONS: Our data indicate that the HeartGuide BP monitor can detect low BP during presyncope and that its measure of SBP change is consistent with that simultaneously obtained through continuous BP monitoring, despite some intrapatient variability. Thus, this device might be useful in determining the hypotensive nature of spontaneous (pre)syncopal symptoms, a possibility that should be verified by field studies.
Articolo in rivista - Articolo scientifico
blood pressure monitoring; finger blood pressure; syncope; wearable (watch-type) monitor;
English
5-ago-2022
2022
11
16
e026420
none
Groppelli, A., Rafanelli, M., Testa, G., Agusto, S., Rivasi, G., Ungar, A., et al. (2022). Feasibility of Blood Pressure Measurement With a Wearable (Watch-Type) Monitor During Impending Syncopal Episodes. JOURNAL OF THE AMERICAN HEART ASSOCIATION. CARDIOVASCULAR AND CEREBROVASCULAR DISEASE, 11(16) [10.1161/JAHA.122.026420].
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/10281/423942
Citazioni
  • Scopus 3
  • ???jsp.display-item.citation.isi??? 3
Social impact