This paper analyzes how to improve the precision of ionoacoustic proton range verification by optimizing the analog signal processing stages with particular emphasis on analog filters. The ionoacoustic technique allows one to spatially detect the proton beam penetration depth/range in a water absorber, with interesting possible applications in real‐time beam monitoring during hadron therapy treatments. The state of the art uses nonoptimized detectors that have low signal quality and thus require a higher total dose, which is not compatible with clinical applications. For these reasons, a comprehensive analysis of acoustic signal bandwidth, signal‐to‐noise‐ratio and noise power/bandwidth will be presented. The correlation between these signal‐quality parameters with maximum achievable proton range measurement precision will be discussed. In particular, the use of an optimized analog filter allows one to decrease the dose required to achieve a given precision by as much as 98.4% compared to a nonoptimized filter approach.

Vallicelli, E., De Matteis, M. (2021). Analog filters design for improving precision in proton sound detectors. JOURNAL OF LOW POWER ELECTRONICS AND APPLICATIONS, 11(1) [10.3390/jlpea11010012].

Analog filters design for improving precision in proton sound detectors

Vallicelli E. A.
;
De Matteis M.
2021

Abstract

This paper analyzes how to improve the precision of ionoacoustic proton range verification by optimizing the analog signal processing stages with particular emphasis on analog filters. The ionoacoustic technique allows one to spatially detect the proton beam penetration depth/range in a water absorber, with interesting possible applications in real‐time beam monitoring during hadron therapy treatments. The state of the art uses nonoptimized detectors that have low signal quality and thus require a higher total dose, which is not compatible with clinical applications. For these reasons, a comprehensive analysis of acoustic signal bandwidth, signal‐to‐noise‐ratio and noise power/bandwidth will be presented. The correlation between these signal‐quality parameters with maximum achievable proton range measurement precision will be discussed. In particular, the use of an optimized analog filter allows one to decrease the dose required to achieve a given precision by as much as 98.4% compared to a nonoptimized filter approach.
Articolo in rivista - Articolo scientifico
Acoustic waves; Analog circuits; Digital circuits; Nuclear imaging; Particle accelerator; Particle beam measurements;
English
2021
11
1
12
none
Vallicelli, E., De Matteis, M. (2021). Analog filters design for improving precision in proton sound detectors. JOURNAL OF LOW POWER ELECTRONICS AND APPLICATIONS, 11(1) [10.3390/jlpea11010012].
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/10281/413545
Citazioni
  • Scopus 5
  • ???jsp.display-item.citation.isi??? 5
Social impact