The high-energy neutron field of the VESUVIO instrument at the ISIS facility has been characterized using the technique of thin-film breakdown counters (TFBC). The technique utilizes neutron-induced fission reactions of natU and 209Bi with detection of fission fragments by TFBCs. Experimentally determined count rates of the fragments are ≈50% higher than those calculated using spectral neutron flux simulated with the MCNPX code. This work is a part of the project to develop ChipIr, a new dedicated facility for the accelerated testing of electronic components and systems for neutron-induced single event effects in the new Target Station 2 at ISIS. The TFBC technique has shown to be applicable for on-line monitoring of the neutron flux in the neutron energy range 1–800 MeV at the position of the device under test (DUT).
Smirnov, A., Pietropaolo, A., Prokofiev, A., Rodionova, E., Frost, C., Ansell, S., et al. (2012). Application of thin-film breakdown counters for characterization of neutron field of the VESUVIO instrument at the ISIS spallation source. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH. SECTION A, ACCELERATORS, SPECTROMETERS, DETECTORS AND ASSOCIATED EQUIPMENT, 687, 14-22 [10.1016/j.nima.2012.05.086].
Application of thin-film breakdown counters for characterization of neutron field of the VESUVIO instrument at the ISIS spallation source
GORINI, GIUSEPPE
2012
Abstract
The high-energy neutron field of the VESUVIO instrument at the ISIS facility has been characterized using the technique of thin-film breakdown counters (TFBC). The technique utilizes neutron-induced fission reactions of natU and 209Bi with detection of fission fragments by TFBCs. Experimentally determined count rates of the fragments are ≈50% higher than those calculated using spectral neutron flux simulated with the MCNPX code. This work is a part of the project to develop ChipIr, a new dedicated facility for the accelerated testing of electronic components and systems for neutron-induced single event effects in the new Target Station 2 at ISIS. The TFBC technique has shown to be applicable for on-line monitoring of the neutron flux in the neutron energy range 1–800 MeV at the position of the device under test (DUT).I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.