Various dosimeters have been tested for assessing absorbed doses with microscopic spatial resolution in targets irradiated by high-flux, synchrotron-generated, low-energy (∼30-300 keV) x-ray microbeams. A MOSFET detector has been used for this study since its radio sensitive element, which is extraordinarily narrow (∼1 μm), suits the main applications of interest, microbeam radiation biology and microbeam radiation therapy (MRT). In MRT, micrometer-wide, centimeter-high, and vertically oriented swaths of tissue are irradiated by arrays of rectangular x-ray microbeams produced by a multislit collimator (MSC). We used MOSFETs to measure the dose distribution, produced by arrays of x-ray microbeams shaped by two different MSCs, in a tissue-equivalent phantom. Doses were measured near the center of the arrays and maximum/minimum (peak/valley) dose ratios (PVDRs) were calculated to determine how variations in heights and in widths of the microbeams influenced this for the therapy, potentially important parameter. Monte Carlo (MC) simulations of the absorbed dose distribution in the phantom were also performed. The results show that when the heights of the irradiated swaths were below those applicable to clinical therapy (<1 mm) the MC simulations produce estimates of PVDRs that are up to a factor of 3 higher than the measured values. For arrays of higher microbeams (i.e., 25 μm×1 cm instead of 25×500 μ m2), this difference between measured and simulated PVDRs becomes less than 50%. Closer agreement was observed between the measured and simulated PVDRs for the Tecomet MSC (current collimator design) than for the Archer MSC. Sources of discrepancies between measured and simulated doses are discussed, of which the energy dependent response of the MOSFET was shown to be among the most important.

Siegbahn, E., Braeuer-Krisch, E., Bravin, A., Nettelbeck, H., Lerch, M., Rosenfeld, A. (2009). MOSFET dosimetry with high spatial resolution in intense synchrotron-generated x-ray microbeams. MEDICAL PHYSICS, 36(4), 1128-1137 [10.1118/1.3081934].

MOSFET dosimetry with high spatial resolution in intense synchrotron-generated x-ray microbeams

Bravin A
Membro del Collaboration Group
;
2009

Abstract

Various dosimeters have been tested for assessing absorbed doses with microscopic spatial resolution in targets irradiated by high-flux, synchrotron-generated, low-energy (∼30-300 keV) x-ray microbeams. A MOSFET detector has been used for this study since its radio sensitive element, which is extraordinarily narrow (∼1 μm), suits the main applications of interest, microbeam radiation biology and microbeam radiation therapy (MRT). In MRT, micrometer-wide, centimeter-high, and vertically oriented swaths of tissue are irradiated by arrays of rectangular x-ray microbeams produced by a multislit collimator (MSC). We used MOSFETs to measure the dose distribution, produced by arrays of x-ray microbeams shaped by two different MSCs, in a tissue-equivalent phantom. Doses were measured near the center of the arrays and maximum/minimum (peak/valley) dose ratios (PVDRs) were calculated to determine how variations in heights and in widths of the microbeams influenced this for the therapy, potentially important parameter. Monte Carlo (MC) simulations of the absorbed dose distribution in the phantom were also performed. The results show that when the heights of the irradiated swaths were below those applicable to clinical therapy (<1 mm) the MC simulations produce estimates of PVDRs that are up to a factor of 3 higher than the measured values. For arrays of higher microbeams (i.e., 25 μm×1 cm instead of 25×500 μ m2), this difference between measured and simulated PVDRs becomes less than 50%. Closer agreement was observed between the measured and simulated PVDRs for the Tecomet MSC (current collimator design) than for the Archer MSC. Sources of discrepancies between measured and simulated doses are discussed, of which the energy dependent response of the MOSFET was shown to be among the most important.
Articolo in rivista - Articolo scientifico
MOSFET dosimetry, high spatial resolution, x-rays
English
2009
36
4
1128
1137
reserved
Siegbahn, E., Braeuer-Krisch, E., Bravin, A., Nettelbeck, H., Lerch, M., Rosenfeld, A. (2009). MOSFET dosimetry with high spatial resolution in intense synchrotron-generated x-ray microbeams. MEDICAL PHYSICS, 36(4), 1128-1137 [10.1118/1.3081934].
File in questo prodotto:
File Dimensione Formato  
siegbahn_medphys_2009.pdf

Solo gestori archivio

Tipologia di allegato: Publisher’s Version (Version of Record, VoR)
Dimensione 371.41 kB
Formato Adobe PDF
371.41 kB Adobe PDF   Visualizza/Apri   Richiedi una copia

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/10281/343118
Citazioni
  • Scopus 39
  • ???jsp.display-item.citation.isi??? 38
Social impact