The evaluation of the image quality of the angiography imaging system used at the Medical Beamline of European Synchrotron Radiation Facility (ESRF) is presented. A two-line 432-pixel germanium detector is utilized in the human coronary angiography program. The performances have been measured with two monochromatic beams, with energies above and below the iodine K-edge. The large-area transfer characteristics (response curve), the spatial detail transfer performance (modulation transfer function (MTF)) and the noise properties (noise power spectrum (NPS)) have been measured and used to compute the frequency-dependent detective quantum efficiency DQE of the system.

Peterzol, A., Bravin, A., Coan, P., Elleaume, H. (2003). Image quality evaluation of the angiography imaging system at the European synchrotron radiation facility. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH. SECTION A, ACCELERATORS, SPECTROMETERS, DETECTORS AND ASSOCIATED EQUIPMENT, 510(1-2), 45-50 [10.1016/S0168-9002(03)01677-2].

Image quality evaluation of the angiography imaging system at the European synchrotron radiation facility

Bravin A
Secondo
Membro del Collaboration Group
;
2003

Abstract

The evaluation of the image quality of the angiography imaging system used at the Medical Beamline of European Synchrotron Radiation Facility (ESRF) is presented. A two-line 432-pixel germanium detector is utilized in the human coronary angiography program. The performances have been measured with two monochromatic beams, with energies above and below the iodine K-edge. The large-area transfer characteristics (response curve), the spatial detail transfer performance (modulation transfer function (MTF)) and the noise properties (noise power spectrum (NPS)) have been measured and used to compute the frequency-dependent detective quantum efficiency DQE of the system.
Articolo in rivista - Articolo scientifico
X-rays, MTF, detector, characterization
English
2003
510
1-2
45
50
reserved
Peterzol, A., Bravin, A., Coan, P., Elleaume, H. (2003). Image quality evaluation of the angiography imaging system at the European synchrotron radiation facility. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH. SECTION A, ACCELERATORS, SPECTROMETERS, DETECTORS AND ASSOCIATED EQUIPMENT, 510(1-2), 45-50 [10.1016/S0168-9002(03)01677-2].
File in questo prodotto:
File Dimensione Formato  
peterzol-nim-angio.pdf

Solo gestori archivio

Tipologia di allegato: Publisher’s Version (Version of Record, VoR)
Dimensione 365.26 kB
Formato Adobe PDF
365.26 kB Adobe PDF   Visualizza/Apri   Richiedi una copia

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/10281/342907
Citazioni
  • Scopus 10
  • ???jsp.display-item.citation.isi??? 10
Social impact