Pain intensity can provide relevant medical information; nowadays, pain measuring is still subjective; therefore, there are efforts to develop devices to measure the intensity of painful stimuli objectively. A review of the scientific literature on pressure pain obtained through algometers is carried out in this work. This literature review aims to present a database of pain pressure thresholds (PPT) related to different body parts so that future research on PPT, using an algometer, could compare their results against this research. The results of these PPT values are presented and grouped according to the body regions where previous researchers recorded their results to where they applied the pressure- trapezius muscles, back muscles, upper extremities, leg muscles, foot area, and muscles of the head. These values can be considered a reference to evaluate new algometers' accuracy and reliability. Finally, the algometers operational parameters range are defined according to all the studied algometers.

Trueba-Perdomo, J., Gasparini, F., Flores Cuautle, J. (2021). Pressure Pain Threshold Values Obtained Through Algometers. REVISTA MEXICANA DE INGENIERÍA BIOMÉDICA, 42(2), 132-148 [10.17488/RMIB.42.2.11].

Pressure Pain Threshold Values Obtained Through Algometers

Gasparini, F
Secondo
;
2021

Abstract

Pain intensity can provide relevant medical information; nowadays, pain measuring is still subjective; therefore, there are efforts to develop devices to measure the intensity of painful stimuli objectively. A review of the scientific literature on pressure pain obtained through algometers is carried out in this work. This literature review aims to present a database of pain pressure thresholds (PPT) related to different body parts so that future research on PPT, using an algometer, could compare their results against this research. The results of these PPT values are presented and grouped according to the body regions where previous researchers recorded their results to where they applied the pressure- trapezius muscles, back muscles, upper extremities, leg muscles, foot area, and muscles of the head. These values can be considered a reference to evaluate new algometers' accuracy and reliability. Finally, the algometers operational parameters range are defined according to all the studied algometers.
Articolo in rivista - Articolo scientifico
Electromechanical algometer; Manual algometer; Pain threshold; Pressure;
English
2021
42
2
132
148
open
Trueba-Perdomo, J., Gasparini, F., Flores Cuautle, J. (2021). Pressure Pain Threshold Values Obtained Through Algometers. REVISTA MEXICANA DE INGENIERÍA BIOMÉDICA, 42(2), 132-148 [10.17488/RMIB.42.2.11].
File in questo prodotto:
File Dimensione Formato  
Pain2021Mexican.pdf

accesso aperto

Descrizione: articolo
Tipologia di allegato: Publisher’s Version (Version of Record, VoR)
Dimensione 498.27 kB
Formato Adobe PDF
498.27 kB Adobe PDF Visualizza/Apri

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.