The first SiPM-based clinical SPECT ring for brain simultaneous imaging in standard MR scanners has been put into operation. Reliable operation of the instrument has been achieved by means of its optimization, such as the introduction of a dual-cooling system (liquid-gas) of the DAQ electronics. The static gantry is composed of 20 MRI-compatible γ-detection modules mounting 5 cm × 10 cm CsI(Tl) scintillators for a total of 1440 pixels (targeting a transaxial FoV of 20 cm). Here we present the results of a preliminary characterization showing 220 kcps max. count rate, 365cps/MBq sensitivity with the multislit-slat collimator, ~17% energy resolution and an average extrinsic spatial resolution of ~8 mm in the trans-axial plane.

Carminati, M., D'Adda, I., Morahan, A., Erlandsson, K., Nagy, K., Nyitrai, Z., et al. (2019). A SiPM-Based Clinical MRI-Compatible SPECT Insert. In 2019 IEEE Nuclear Science Symposium and Medical Imaging Conference, NSS/MIC 2019 (pp.1-3). Institute of Electrical and Electronics Engineers Inc. [10.1109/NSS/MIC42101.2019.9060025].

A SiPM-Based Clinical MRI-Compatible SPECT Insert

Moresco R.;Savi A.;
2019

Abstract

The first SiPM-based clinical SPECT ring for brain simultaneous imaging in standard MR scanners has been put into operation. Reliable operation of the instrument has been achieved by means of its optimization, such as the introduction of a dual-cooling system (liquid-gas) of the DAQ electronics. The static gantry is composed of 20 MRI-compatible γ-detection modules mounting 5 cm × 10 cm CsI(Tl) scintillators for a total of 1440 pixels (targeting a transaxial FoV of 20 cm). Here we present the results of a preliminary characterization showing 220 kcps max. count rate, 365cps/MBq sensitivity with the multislit-slat collimator, ~17% energy resolution and an average extrinsic spatial resolution of ~8 mm in the trans-axial plane.
paper
Cesium iodide, Electronic cooling, Liquefied gases
English
2019 IEEE Nuclear Science Symposium and Medical Imaging Conference, NSS/MIC 2019
2019
2019 IEEE Nuclear Science Symposium and Medical Imaging Conference, NSS/MIC 2019
9781728141640
2019
1
3
9060025
none
Carminati, M., D'Adda, I., Morahan, A., Erlandsson, K., Nagy, K., Nyitrai, Z., et al. (2019). A SiPM-Based Clinical MRI-Compatible SPECT Insert. In 2019 IEEE Nuclear Science Symposium and Medical Imaging Conference, NSS/MIC 2019 (pp.1-3). Institute of Electrical and Electronics Engineers Inc. [10.1109/NSS/MIC42101.2019.9060025].
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/10281/281596
Citazioni
  • Scopus 0
  • ???jsp.display-item.citation.isi??? 1
Social impact