Brachial cuff blood pressure (BP) is clinically important, but may be an inaccurate substitute for central BP. Many non-invasive devices have been developed that purport to estimate central BP from peripheral artery sites, yet with no standardized guidelines; the accuracy testing of these new devices has not been undertaken in a uniform fashion with comparable protocols. This is an abridged paper describing the recommendations reached by an international task force convened to identify issues that need to be addressed and reach consensus relating to methods for assessing and reporting the accuracy (validation) of central BP devices. The recommendations are endorsed by the Association for Research into Arterial Structure and Physiology (ARTERY) Society, as well as the European Society of Hypertension (ESH) Working Group on Arterial Structure and Function, and the ESH Working Group on Blood Pressure Monitoring and Cardiovascular Variability. Researchers interested in validating central BP monitors should read the full version of the statement.

Sharman, J., Avolio, A., Baulmann, J., Benetos, A., Blacher, J., Blizzard, C., et al. (2017). Validation of non-invasive central blood pressure devices: Artery society task force (abridged) consensus statement on protocol standardization. ARTERY RESEARCH, 20, 35-43 [10.1016/j.artres.2017.11.001].

Validation of non-invasive central blood pressure devices: Artery society task force (abridged) consensus statement on protocol standardization

Parati G.;
2017

Abstract

Brachial cuff blood pressure (BP) is clinically important, but may be an inaccurate substitute for central BP. Many non-invasive devices have been developed that purport to estimate central BP from peripheral artery sites, yet with no standardized guidelines; the accuracy testing of these new devices has not been undertaken in a uniform fashion with comparable protocols. This is an abridged paper describing the recommendations reached by an international task force convened to identify issues that need to be addressed and reach consensus relating to methods for assessing and reporting the accuracy (validation) of central BP devices. The recommendations are endorsed by the Association for Research into Arterial Structure and Physiology (ARTERY) Society, as well as the European Society of Hypertension (ESH) Working Group on Arterial Structure and Function, and the ESH Working Group on Blood Pressure Monitoring and Cardiovascular Variability. Researchers interested in validating central BP monitors should read the full version of the statement.
Articolo in rivista - Articolo scientifico
Aorta; Central blood pressure; Diagnostic equipment; Guideline; Hypertension;
Aorta, Central blood pressure, Diagnostic equipment, Guideline, Hypertension
English
2017
20
35
43
open
Sharman, J., Avolio, A., Baulmann, J., Benetos, A., Blacher, J., Blizzard, C., et al. (2017). Validation of non-invasive central blood pressure devices: Artery society task force (abridged) consensus statement on protocol standardization. ARTERY RESEARCH, 20, 35-43 [10.1016/j.artres.2017.11.001].
File in questo prodotto:
File Dimensione Formato  
125924969.pdf

accesso aperto

Tipologia di allegato: Publisher’s Version (Version of Record, VoR)
Licenza: Creative Commons
Dimensione 868.66 kB
Formato Adobe PDF
868.66 kB Adobe PDF Visualizza/Apri

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/10281/280137
Citazioni
  • Scopus 7
  • ???jsp.display-item.citation.isi??? 7
Social impact