The ICARUS T600 LAr TPC is the far detector of the Short Baseline Program at FNAL. As it will have to work at shallow depth in the Booster Neutrino Beam, a large cosmic rays background (∼ 11 kHz) will be present. To reduce it, precise timing information is needed from the new light detection system, based on 360 large area photomultipliers. For precise time measurements a calibration system based on a fast laser diode and a system based on one optical switch, several 01 × 1 fused fiber splitters, ultra-high vacuum optical feedthroughs and multimode optical patchcords up to 20 m long, to distribute the laser pulses to each single PMT, was designed. The time evolution of the PMTs' gain/timing and possibly their initial calibrations at a time t0 will be done by using this system. The expected time resolution of this calibration system will be around 100 ps. The laboratory tests needed to set up the system are reported.

Bonesini, M., Benocci, R., Bertoni, R., Falcone, A., Mazza, R., Torti, M., et al. (2020). The laser diode calibration system of the ICARUS T600 detector at FNAL. Intervento presentato a: 15th Topical Seminar on Innovative Particle and Radiation Detectors (IPRD19), Siena, Italy [10.1088/1748-0221/15/05/C05042].

The laser diode calibration system of the ICARUS T600 detector at FNAL

Bonesini M.;Benocci R.;Falcone A.;Torti M.;
2020

Abstract

The ICARUS T600 LAr TPC is the far detector of the Short Baseline Program at FNAL. As it will have to work at shallow depth in the Booster Neutrino Beam, a large cosmic rays background (∼ 11 kHz) will be present. To reduce it, precise timing information is needed from the new light detection system, based on 360 large area photomultipliers. For precise time measurements a calibration system based on a fast laser diode and a system based on one optical switch, several 01 × 1 fused fiber splitters, ultra-high vacuum optical feedthroughs and multimode optical patchcords up to 20 m long, to distribute the laser pulses to each single PMT, was designed. The time evolution of the PMTs' gain/timing and possibly their initial calibrations at a time t0 will be done by using this system. The expected time resolution of this calibration system will be around 100 ps. The laboratory tests needed to set up the system are reported.
paper
Detector alignment and calibration methods (lasers
Neutrino detectors
particle-beams)
sources
Time projection Chambers (TPC)
English
15th Topical Seminar on Innovative Particle and Radiation Detectors (IPRD19)
2019
2020
15
5
C05042
C05042
C05042
reserved
Bonesini, M., Benocci, R., Bertoni, R., Falcone, A., Mazza, R., Torti, M., et al. (2020). The laser diode calibration system of the ICARUS T600 detector at FNAL. Intervento presentato a: 15th Topical Seminar on Innovative Particle and Radiation Detectors (IPRD19), Siena, Italy [10.1088/1748-0221/15/05/C05042].
File in questo prodotto:
File Dimensione Formato  
2019 Siena_2.pdf

Solo gestori archivio

Tipologia di allegato: Author’s Accepted Manuscript, AAM (Post-print)
Dimensione 3.11 MB
Formato Adobe PDF
3.11 MB Adobe PDF   Visualizza/Apri   Richiedi una copia

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.