Intermittent noninvasive blood pressure monitoring is becoming increasingly popular because of the belief that a daytime blood pressure profile can provide a better clinical evaluation of hypertension than that provided by casual blood pressure measurements. This approach has potential limitations, however, one of which is that the cuff inflations permitting blood pressure to be repeatedly measured may induce an alerting reaction and a pressor response in the patients and lead to an overestimation of their daytime blood pressure. Blood pressure in 22 subjects was invasively recorded for 24 hours by the Oxford method. During the day of the recording blood pressure was also measured by a noninvasive device (Vita-Stat 901), which had its cuff applied to the opposite arm from which the intra-arterial signal was derived. For 2 hours the device provided automatic cuff inflations at 10-minute intervals. For another 2 hours it was programmed to provide cuff inflations only following patients’ commands, also at 10-minute intervals. Analysis of the intra-arterial blood pressure trace during the periods preceding and following the automatically or semiautomatically induced cuff inflations showed that these procedures. © 1985, American Heart Association, Inc.

Parati, G., Pomidossi, G., Casadei, R., Mancia, G. (1985). Lack of alerting reactions to intermittent cuff inflations during noninvasive blood pressure monitoring. HYPERTENSION, 7(4), 597-601 [10.1161/01.HYP.7.4.597].

Lack of alerting reactions to intermittent cuff inflations during noninvasive blood pressure monitoring

Parati G.;Mancia G.
1985

Abstract

Intermittent noninvasive blood pressure monitoring is becoming increasingly popular because of the belief that a daytime blood pressure profile can provide a better clinical evaluation of hypertension than that provided by casual blood pressure measurements. This approach has potential limitations, however, one of which is that the cuff inflations permitting blood pressure to be repeatedly measured may induce an alerting reaction and a pressor response in the patients and lead to an overestimation of their daytime blood pressure. Blood pressure in 22 subjects was invasively recorded for 24 hours by the Oxford method. During the day of the recording blood pressure was also measured by a noninvasive device (Vita-Stat 901), which had its cuff applied to the opposite arm from which the intra-arterial signal was derived. For 2 hours the device provided automatic cuff inflations at 10-minute intervals. For another 2 hours it was programmed to provide cuff inflations only following patients’ commands, also at 10-minute intervals. Analysis of the intra-arterial blood pressure trace during the periods preceding and following the automatically or semiautomatically induced cuff inflations showed that these procedures. © 1985, American Heart Association, Inc.
Articolo in rivista - Articolo scientifico
Alerting responses; Blood pressure; Hypertension; Noninvasive blood pressure monitoring; Stress; Adult; Aged; Blood Pressure Determination; Female; Heart Rate; Humans; Male; Middle Aged; Blood Pressure; Monitoring, Physiologic
English
1985
7
4
597
601
none
Parati, G., Pomidossi, G., Casadei, R., Mancia, G. (1985). Lack of alerting reactions to intermittent cuff inflations during noninvasive blood pressure monitoring. HYPERTENSION, 7(4), 597-601 [10.1161/01.HYP.7.4.597].
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/10281/273538
Citazioni
  • Scopus 132
  • ???jsp.display-item.citation.isi??? 141
Social impact