A De Pangher-type long counter was constructed at Peking University for neutron fluence measurement. The responses and effective centers of the long counter were calculated using the Monte Carlo method. The variations with neutron energy of the position of the effective center, calculated by the Monte Carlo code, was experimentally validated with a 241AmBe neutron source and several mono-energetic neutron sources ranging from 100 keV to 6 MeV. Long counter calibration was performed using a radionuclide source from Peking University, and neutron fluence measured by the calibrated long counter was successfully compared to values determined with reference instruments (two recoil proton counters and a 238U fission chamber). The relative deviations were lower than 6% along the whole energy range. The calibrated long counter was thus successfully applied in the calibration of a Bonner sphere spectrometer.

Hu, Z., Peng, X., Chen, Z., Du, T., Ge, L., Yuan, X., et al. (2018). Experimental characterization of a long counter for neutron fluence measurement. RADIATION MEASUREMENTS, 119, 16-21 [10.1016/j.radmeas.2018.08.017].

Experimental characterization of a long counter for neutron fluence measurement

Hu Z. M.;Gorini G.
Penultimo
;
2018

Abstract

A De Pangher-type long counter was constructed at Peking University for neutron fluence measurement. The responses and effective centers of the long counter were calculated using the Monte Carlo method. The variations with neutron energy of the position of the effective center, calculated by the Monte Carlo code, was experimentally validated with a 241AmBe neutron source and several mono-energetic neutron sources ranging from 100 keV to 6 MeV. Long counter calibration was performed using a radionuclide source from Peking University, and neutron fluence measured by the calibrated long counter was successfully compared to values determined with reference instruments (two recoil proton counters and a 238U fission chamber). The relative deviations were lower than 6% along the whole energy range. The calibrated long counter was thus successfully applied in the calibration of a Bonner sphere spectrometer.
Articolo in rivista - Articolo scientifico
Calibration; Effective center; Long counter; Response function;
Calibration; Effective center; Long counter; Response function
English
2018
119
16
21
none
Hu, Z., Peng, X., Chen, Z., Du, T., Ge, L., Yuan, X., et al. (2018). Experimental characterization of a long counter for neutron fluence measurement. RADIATION MEASUREMENTS, 119, 16-21 [10.1016/j.radmeas.2018.08.017].
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/10281/270856
Citazioni
  • Scopus 3
  • ???jsp.display-item.citation.isi??? 3
Social impact