The Lancet Commission on Hypertension identified that a key action to address the worldwide burden of high blood pressure (BP) was to improve the quality of BP measurements by using BP devices that have been validated for accuracy. Currently, there are over 3 000 commercially available BP devices, but many do not have published data on accuracy testing according to established scientific standards. This problem is enabled through weak or absent regulations that allow clearance of devices for commercial use without formal validation. In addition, new BP technologies have emerged (e.g. cuffless sensors) for which there is no scientific consensus regarding BP measurement accuracy standards. Altogether, these issues contribute to the widespread availability of clinic and home BP devices with limited or uncertain accuracy, leading to inappropriate hypertension diagnosis, management and drug treatment on a global scale. The most significant problems relating to the accuracy of BP devices can be resolved by the regulatory requirement for mandatory independent validation of BP devices according to the universally-accepted International Organization for Standardization Standard. This is a primary recommendation for which there is an urgent international need. Other key recommendations are development of validation standards specifically for new BP technologies and online lists of accurate devices that are accessible to consumers and health professionals. Recommendations are aligned with WHO policies on medical devices and universal healthcare. Adherence to recommendations would increase the global availability of accurate BP devices and result in better diagnosis and treatment of hypertension, thus decreasing the worldwide burden from high BP.

Sharman, J., O'Brien, E., Alpert, B., Schutte, A., Delles, C., Olsen, M., et al. (2020). Lancet Commission on Hypertension Group position statement on the global improvement of accuracy standards for devices that measure blood pressure [Declaración de posición del Grupo de la Comisión Lancet de Hipertensión con respecto a la mejora mundial de las normas de exactitud para los dispositivos de medición de la presión arterial]. REVISTA PANAMERICANA DE SALUD PUBLICA, 44 [10.26633/RPSP.2020.21].

Lancet Commission on Hypertension Group position statement on the global improvement of accuracy standards for devices that measure blood pressure [Declaración de posición del Grupo de la Comisión Lancet de Hipertensión con respecto a la mejora mundial de las normas de exactitud para los dispositivos de medición de la presión arterial]

Parati, Gianfranco;
2020

Abstract

The Lancet Commission on Hypertension identified that a key action to address the worldwide burden of high blood pressure (BP) was to improve the quality of BP measurements by using BP devices that have been validated for accuracy. Currently, there are over 3 000 commercially available BP devices, but many do not have published data on accuracy testing according to established scientific standards. This problem is enabled through weak or absent regulations that allow clearance of devices for commercial use without formal validation. In addition, new BP technologies have emerged (e.g. cuffless sensors) for which there is no scientific consensus regarding BP measurement accuracy standards. Altogether, these issues contribute to the widespread availability of clinic and home BP devices with limited or uncertain accuracy, leading to inappropriate hypertension diagnosis, management and drug treatment on a global scale. The most significant problems relating to the accuracy of BP devices can be resolved by the regulatory requirement for mandatory independent validation of BP devices according to the universally-accepted International Organization for Standardization Standard. This is a primary recommendation for which there is an urgent international need. Other key recommendations are development of validation standards specifically for new BP technologies and online lists of accurate devices that are accessible to consumers and health professionals. Recommendations are aligned with WHO policies on medical devices and universal healthcare. Adherence to recommendations would increase the global availability of accurate BP devices and result in better diagnosis and treatment of hypertension, thus decreasing the worldwide burden from high BP.
Articolo in rivista - Articolo scientifico
Biomedical technology; diagnostic equipment; international health; reference standard
Spanish; Castilian
2020
44
e21
none
Sharman, J., O'Brien, E., Alpert, B., Schutte, A., Delles, C., Olsen, M., et al. (2020). Lancet Commission on Hypertension Group position statement on the global improvement of accuracy standards for devices that measure blood pressure [Declaración de posición del Grupo de la Comisión Lancet de Hipertensión con respecto a la mejora mundial de las normas de exactitud para los dispositivos de medición de la presión arterial]. REVISTA PANAMERICANA DE SALUD PUBLICA, 44 [10.26633/RPSP.2020.21].
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/10281/265185
Citazioni
  • Scopus 3
  • ???jsp.display-item.citation.isi??? 0
Social impact