A slit-lamp biomicroscope Visionix VX75 has been equipped with a high- resolution digital sensor. A specular reflection technique at an angular magnifica- tion of 36 performed by the slit-lamp biomicroscope is used to develop a proce- dure to (i) measure the thickness of the human cornea by measuring the distance between the two reflections of its anterior and posterior surfaces and (ii) capture suitable images for morphometric analyses of the corneal endothelium’s cell mosaic. The examples of morphometric analysis of these images are reported. The biases due to the dioptric power of the anterior surface of the cornea, the oblique obser- vation, and the asymmetry of the digital biomicroscope are discussed. These biases can be corrected by a specific calibration.

Tavazzi, S., Parodi, A., Colciago, S., Nigrotti, G., Borghesi, S., Zeri, F. (2019). Corneal Pachymetry and Endothelial Microscopy by Slit-Lamp. In A. Nowinska (a cura di), Novel Diagnostic Methods in Ophthalmology (pp. 1-14). IntechOpen [10.5772/intechopen.85037].

Corneal Pachymetry and Endothelial Microscopy by Slit-Lamp

Tavazzi, S
;
Nigrotti, G;Borghesi, S;Zeri, F
2019

Abstract

A slit-lamp biomicroscope Visionix VX75 has been equipped with a high- resolution digital sensor. A specular reflection technique at an angular magnifica- tion of 36 performed by the slit-lamp biomicroscope is used to develop a proce- dure to (i) measure the thickness of the human cornea by measuring the distance between the two reflections of its anterior and posterior surfaces and (ii) capture suitable images for morphometric analyses of the corneal endothelium’s cell mosaic. The examples of morphometric analysis of these images are reported. The biases due to the dioptric power of the anterior surface of the cornea, the oblique obser- vation, and the asymmetry of the digital biomicroscope are discussed. These biases can be corrected by a specific calibration.
Capitolo o saggio
slit-lamp biomicroscope, pachymetry, endothelial microscopy, specular reflection, morphometric analysis
English
Novel Diagnostic Methods in Ophthalmology
Nowinska, A
2019
2019
978-1-83880-312-4
IntechOpen
1
14
Tavazzi, S., Parodi, A., Colciago, S., Nigrotti, G., Borghesi, S., Zeri, F. (2019). Corneal Pachymetry and Endothelial Microscopy by Slit-Lamp. In A. Nowinska (a cura di), Novel Diagnostic Methods in Ophthalmology (pp. 1-14). IntechOpen [10.5772/intechopen.85037].
open
File in questo prodotto:
File Dimensione Formato  
corneal pachymetry.pdf

accesso aperto

Descrizione: Articolo
Tipologia di allegato: Publisher’s Version (Version of Record, VoR)
Dimensione 2.71 MB
Formato Adobe PDF
2.71 MB Adobe PDF Visualizza/Apri

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/10281/260214
Citazioni
  • Scopus ND
  • ???jsp.display-item.citation.isi??? ND
Social impact