Objectives In this study, we compared left ventricular (LV) segmental strain measurements obtained with different ultrasound machines and post-processing software packages. Background Global longitudinal strain (GLS) has proven to be a reproducible and valuable tool in clinical practice. Data about the reproducibility and intervendor differences of segmental strain measurements, however, are missing. Methods We included 63 volunteers with cardiac magnetic resonance–proven infarct scar with segmental LV function ranging from normal to severely impaired. Each subject was examined within 2 h by a single expert sonographer with machines from multiple vendors. All 3 apical views were acquired twice to determine the test-retest and the intervendor variability. Segmental longitudinal peak systolic, end-systolic, and post-systolic strain were measured using 7 vendor-specific systems (Hitachi, Tokyo, Japan; Esaote, Florence, Italy; GE Vingmed Ultrasound, Horten, Norway; Philips, Andover, Massachusetts; Samsung, Seoul, South Korea; Siemens, Mountain View, California; and Toshiba, Otawara, Japan) and 2 independent software packages (Epsilon, Ann Arbor, Michigan; and TOMTEC, Unterschleissheim, Germany) and compared among vendors. Results Image quality and tracking feasibility differed among vendors (analysis of variance, p < 0.05). The absolute test-retest difference ranged from 2.5% to 4.9% for peak systolic, 2.6% to 5.0% for end-systolic, and 2.5% to 5.0% for post-systolic strain. The average segmental strain values varied significantly between vendors (up to 4.5%). Segmental strain parameters from each vendor correlated well with the mean of all vendors (r2 range 0.58 to 0.81) but showed very different ranges of values. Bias and limits of agreement were up to −4.6 ± 7.5%. Conclusions In contrast to GLS, LV segmental longitudinal strain measurements have a higher variability on top of the known intervendor bias. The fidelity of different software to follow segmental function varies considerably. We conclude that single segmental strain values should be used with caution in the clinic. Segmental strain pattern analysis might be a more robust alternative.

Mirea, O., Pagourelias, E., Duchenne, J., Bogaert, J., Thomas, J., Badano, L., et al. (2018). Variability and Reproducibility of Segmental Longitudinal Strain Measurement: A Report From the EACVI-ASE Strain Standardization Task Force. JACC. CARDIOVASCULAR IMAGING, 11(1), 15-24 [10.1016/j.jcmg.2017.01.027].

Variability and Reproducibility of Segmental Longitudinal Strain Measurement: A Report From the EACVI-ASE Strain Standardization Task Force

Badano L.;
2018

Abstract

Objectives In this study, we compared left ventricular (LV) segmental strain measurements obtained with different ultrasound machines and post-processing software packages. Background Global longitudinal strain (GLS) has proven to be a reproducible and valuable tool in clinical practice. Data about the reproducibility and intervendor differences of segmental strain measurements, however, are missing. Methods We included 63 volunteers with cardiac magnetic resonance–proven infarct scar with segmental LV function ranging from normal to severely impaired. Each subject was examined within 2 h by a single expert sonographer with machines from multiple vendors. All 3 apical views were acquired twice to determine the test-retest and the intervendor variability. Segmental longitudinal peak systolic, end-systolic, and post-systolic strain were measured using 7 vendor-specific systems (Hitachi, Tokyo, Japan; Esaote, Florence, Italy; GE Vingmed Ultrasound, Horten, Norway; Philips, Andover, Massachusetts; Samsung, Seoul, South Korea; Siemens, Mountain View, California; and Toshiba, Otawara, Japan) and 2 independent software packages (Epsilon, Ann Arbor, Michigan; and TOMTEC, Unterschleissheim, Germany) and compared among vendors. Results Image quality and tracking feasibility differed among vendors (analysis of variance, p < 0.05). The absolute test-retest difference ranged from 2.5% to 4.9% for peak systolic, 2.6% to 5.0% for end-systolic, and 2.5% to 5.0% for post-systolic strain. The average segmental strain values varied significantly between vendors (up to 4.5%). Segmental strain parameters from each vendor correlated well with the mean of all vendors (r2 range 0.58 to 0.81) but showed very different ranges of values. Bias and limits of agreement were up to −4.6 ± 7.5%. Conclusions In contrast to GLS, LV segmental longitudinal strain measurements have a higher variability on top of the known intervendor bias. The fidelity of different software to follow segmental function varies considerably. We conclude that single segmental strain values should be used with caution in the clinic. Segmental strain pattern analysis might be a more robust alternative.
Articolo in rivista - Articolo scientifico
intervendor bias; segmental strain; Biomechanical Phenomena; Case-Control Studies; Echocardiography, Doppler; Humans; Image Interpretation, Computer-Assisted; Magnetic Resonance Imaging; Myocardial Infarction; Observer Variation; Predictive Value of Tests; Reproducibility of Results; Software; Stress, Mechanical; Ventricular Dysfunction, Left; Myocardial Contraction; Ventricular Function, Left
English
2018
11
1
15
24
reserved
Mirea, O., Pagourelias, E., Duchenne, J., Bogaert, J., Thomas, J., Badano, L., et al. (2018). Variability and Reproducibility of Segmental Longitudinal Strain Measurement: A Report From the EACVI-ASE Strain Standardization Task Force. JACC. CARDIOVASCULAR IMAGING, 11(1), 15-24 [10.1016/j.jcmg.2017.01.027].
File in questo prodotto:
File Dimensione Formato  
Variability JACC Imaging.pdf

Solo gestori archivio

Tipologia di allegato: Publisher’s Version (Version of Record, VoR)
Dimensione 1.04 MB
Formato Adobe PDF
1.04 MB Adobe PDF   Visualizza/Apri   Richiedi una copia

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/10281/257356
Citazioni
  • Scopus 151
  • ???jsp.display-item.citation.isi??? 137
Social impact