Background: Intervendor consistency of left ventricular (LV) volume measurements using three-dimensional transthoracic echocardiography (3DTTE) has never been reported. Accordingly, we designed a prospective study to (1) compare head-to-head the accuracy of three three-dimensional echocardiography (3DE) systems in measuring LV volumes and ejection fraction (EF) against cardiac magnetic resonance (CMR); (2) assess the intervendor variability of LV volumes and EF; and (3) compare the accuracy of fully automated versus semiautomated (i.e., manually corrected) methods of LV endocardial delineation against CMR.Methods: We studied 92 patients (64% males, 52 years [95% CI, 20-83]) with a wide range of end-diastolic volumes (from 87 to 446 mL) and EFs (from 16% to 77%) using three different 3DE platforms (iE33; VividE9; Acuson SC2000) during the same echo study. CMR was performed within 3 6 5 hours from the 3DE study in 35 patients.Results: LV volumes provided by the three 3DE systems correlated with CMR volumes: end-diastolic volume (iE33: R-2 = 0.93; E9: R-2 = 0.94; SC2000: R-2 = 0.94), end-systolic volume (iE33: R-2 = 0.93; E9: R-2 = 0.95; SC2000: R-2 = 0.94), and EF (iE33: R-2 = 0.79; E9: R-2 = 0.80; SC2000: R-2 = 0.77). In the 92 patients studied, LV volumes and EFs measured with the three systems were similar. Use of fully automated endocardial border detection algorithms significantly underestimated LV volumes, and the degree of underestimation was higher with larger LV volumes.Conclusions: LV volumes and EFs measured with the three 3DE systems are consistent. Fully automated algorithms underestimated LV volumes. Our findings may help in the clinical interpretation of LV parameters obtained using different 3DE systems and encourage the clinical use of 3DTTE.

Muraru, D., Cecchetto, A., Cucchini, U., Zhou, X., Lang, R., Romeo, G., et al. (2018). Intervendor Consistency and Accuracy of Left Ventricular Volume Measurements Using Three-Dimensional Echocardiography. JOURNAL OF THE AMERICAN SOCIETY OF ECHOCARDIOGRAPHY, 31(2), 158-168.e1 [10.1016/j.echo.2017.10.010].

Intervendor Consistency and Accuracy of Left Ventricular Volume Measurements Using Three-Dimensional Echocardiography

Muraru D.;Badano L.
2018

Abstract

Background: Intervendor consistency of left ventricular (LV) volume measurements using three-dimensional transthoracic echocardiography (3DTTE) has never been reported. Accordingly, we designed a prospective study to (1) compare head-to-head the accuracy of three three-dimensional echocardiography (3DE) systems in measuring LV volumes and ejection fraction (EF) against cardiac magnetic resonance (CMR); (2) assess the intervendor variability of LV volumes and EF; and (3) compare the accuracy of fully automated versus semiautomated (i.e., manually corrected) methods of LV endocardial delineation against CMR.Methods: We studied 92 patients (64% males, 52 years [95% CI, 20-83]) with a wide range of end-diastolic volumes (from 87 to 446 mL) and EFs (from 16% to 77%) using three different 3DE platforms (iE33; VividE9; Acuson SC2000) during the same echo study. CMR was performed within 3 6 5 hours from the 3DE study in 35 patients.Results: LV volumes provided by the three 3DE systems correlated with CMR volumes: end-diastolic volume (iE33: R-2 = 0.93; E9: R-2 = 0.94; SC2000: R-2 = 0.94), end-systolic volume (iE33: R-2 = 0.93; E9: R-2 = 0.95; SC2000: R-2 = 0.94), and EF (iE33: R-2 = 0.79; E9: R-2 = 0.80; SC2000: R-2 = 0.77). In the 92 patients studied, LV volumes and EFs measured with the three systems were similar. Use of fully automated endocardial border detection algorithms significantly underestimated LV volumes, and the degree of underestimation was higher with larger LV volumes.Conclusions: LV volumes and EFs measured with the three 3DE systems are consistent. Fully automated algorithms underestimated LV volumes. Our findings may help in the clinical interpretation of LV parameters obtained using different 3DE systems and encourage the clinical use of 3DTTE.
Articolo in rivista - Articolo scientifico
Cardiac magnetic resonance; Ejection fraction; Fully automated; Left ventricle; Repeatability; Reproducibility; Three-dimensional echocardiography; Volumes
English
2018
31
2
158
168.e1
none
Muraru, D., Cecchetto, A., Cucchini, U., Zhou, X., Lang, R., Romeo, G., et al. (2018). Intervendor Consistency and Accuracy of Left Ventricular Volume Measurements Using Three-Dimensional Echocardiography. JOURNAL OF THE AMERICAN SOCIETY OF ECHOCARDIOGRAPHY, 31(2), 158-168.e1 [10.1016/j.echo.2017.10.010].
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/10281/246915
Citazioni
  • Scopus 32
  • ???jsp.display-item.citation.isi??? 26
Social impact