This paper describes the design and performance of a computer controlled emulsion Plate Changer for the automatic placement and removal of nuclear emulsion films for the European Scanning System microscopes. The Plate Changer is used for mass scanning and measurement of the emulsions of the OPERA neutrino oscillation experiment at the Gran Sasso lab on the CNGS neutrino beam. Unlike other systems it works with both dry and oil objectives. The film changing takes less than 20 s and the accuracy on the positioning of the emulsion films is about 10μm. The final accuracy in retrieving track coordinates after fiducial marks measurement is better than 1μm. © 2013 Elsevier B.V.

Balestra, S., Bertolin, A., Bozza, C., Calligola, P., Cerroni, R., D'Ambrosio, N., et al. (2013). A fast automatic plate changer for the analysis of nuclear emulsions. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH. SECTION A, ACCELERATORS, SPECTROMETERS, DETECTORS AND ASSOCIATED EQUIPMENT, 716, 96-100 [10.1016/j.nima.2013.03.021].

A fast automatic plate changer for the analysis of nuclear emulsions

Tenti, M.;
2013

Abstract

This paper describes the design and performance of a computer controlled emulsion Plate Changer for the automatic placement and removal of nuclear emulsion films for the European Scanning System microscopes. The Plate Changer is used for mass scanning and measurement of the emulsions of the OPERA neutrino oscillation experiment at the Gran Sasso lab on the CNGS neutrino beam. Unlike other systems it works with both dry and oil objectives. The film changing takes less than 20 s and the accuracy on the positioning of the emulsion films is about 10μm. The final accuracy in retrieving track coordinates after fiducial marks measurement is better than 1μm. © 2013 Elsevier B.V.
Articolo in rivista - Articolo scientifico
Emulsion plate changer; European Scanning System; Neutrinos; Nuclear emulsions; OPERA experiment; Nuclear and High Energy Physics; Instrumentation
English
2013
716
96
100
none
Balestra, S., Bertolin, A., Bozza, C., Calligola, P., Cerroni, R., D'Ambrosio, N., et al. (2013). A fast automatic plate changer for the analysis of nuclear emulsions. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH. SECTION A, ACCELERATORS, SPECTROMETERS, DETECTORS AND ASSOCIATED EQUIPMENT, 716, 96-100 [10.1016/j.nima.2013.03.021].
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/10281/212950
Citazioni
  • Scopus 3
  • ???jsp.display-item.citation.isi??? 3
Social impact