BACKGROUND: This multicenter, prospective study evaluated the determinants of zero-fluoroscopy (ZFL) ablation of supraventricular tachycardias.METHODS AND RESULTS: Four hundred thirty patients (215 male, 55.4 +/- 22.1 years) with indication to electrophysiological study or ablation of supraventricular tachycardias were enrolled. All participating physicians agreed to follow the as low as reasonably achievable policy. A procedure was defined as ZFL when no fluoroscopy was used. The total fluoroscopy time inversely correlated to the number of procedures previously performed by each operator since study start (r=-0.112; P=0.02). Two hundred eighty-nine procedures (67.2%) were ZFL; multivariable analysis identified as predictors of ZFL: procedure after the 30th for each operator, compared with procedures up to the ninth (P=0.011; hazard ratio, 3.49; 95% confidence interval [ CI], 1.79-6.80); the type of arrhythmia (P=0.031; electrophysiological study and atrioventricular nodal reentry tachycardia ablation having the highest probability of ZFL; hazard ratio, 6.87; 95% CI, 2.08-22.7 and hazard ratio, 2.02; 95% CI, 1.04-3.91, respectively); the operator's (P=0.002) and patient's age (P=0.009). Among operators, achievement of ZFL varied from 0% to 100%; 8 (22.8%) operators achieved ZFL in <25% of their procedures; 17 (48.6%) operators achieved ZFL in >75% of their procedures. The probability of ZFL increased by 2.8% (hazard ratio, 0.98; 95% CI, 0.97-0.99) as patient's age decreased by 1 year. Acute procedural success was obtained in all cases.CONCLUSIONS: The use of 3-dimensional mapping system completely avoided the use of fluoroscopy in most cases, with very low fluoroscopy time in the remaining and high safety and effectiveness profiles. Achievement of ZFL was predicted by the type of arrhythmia, operator's experience, and patient's age

Pani, A., Belotti, G., Bonanno, C., Bongiorni, M., Bottoni, N., Brambilla, R., et al. (2018). Predictors of Zero X-Ray Ablation for Supraventricular Tachycardias in a Nationwide Multicenter Experience. CIRCULATION. ARRHYTHMIA AND ELECTROPHYSIOLOGY, 11(3) [10.1161/CIRCEP.117.005592].

Predictors of Zero X-Ray Ablation for Supraventricular Tachycardias in a Nationwide Multicenter Experience

de Vito, G;
2018

Abstract

BACKGROUND: This multicenter, prospective study evaluated the determinants of zero-fluoroscopy (ZFL) ablation of supraventricular tachycardias.METHODS AND RESULTS: Four hundred thirty patients (215 male, 55.4 +/- 22.1 years) with indication to electrophysiological study or ablation of supraventricular tachycardias were enrolled. All participating physicians agreed to follow the as low as reasonably achievable policy. A procedure was defined as ZFL when no fluoroscopy was used. The total fluoroscopy time inversely correlated to the number of procedures previously performed by each operator since study start (r=-0.112; P=0.02). Two hundred eighty-nine procedures (67.2%) were ZFL; multivariable analysis identified as predictors of ZFL: procedure after the 30th for each operator, compared with procedures up to the ninth (P=0.011; hazard ratio, 3.49; 95% confidence interval [ CI], 1.79-6.80); the type of arrhythmia (P=0.031; electrophysiological study and atrioventricular nodal reentry tachycardia ablation having the highest probability of ZFL; hazard ratio, 6.87; 95% CI, 2.08-22.7 and hazard ratio, 2.02; 95% CI, 1.04-3.91, respectively); the operator's (P=0.002) and patient's age (P=0.009). Among operators, achievement of ZFL varied from 0% to 100%; 8 (22.8%) operators achieved ZFL in <25% of their procedures; 17 (48.6%) operators achieved ZFL in >75% of their procedures. The probability of ZFL increased by 2.8% (hazard ratio, 0.98; 95% CI, 0.97-0.99) as patient's age decreased by 1 year. Acute procedural success was obtained in all cases.CONCLUSIONS: The use of 3-dimensional mapping system completely avoided the use of fluoroscopy in most cases, with very low fluoroscopy time in the remaining and high safety and effectiveness profiles. Achievement of ZFL was predicted by the type of arrhythmia, operator's experience, and patient's age
Articolo in rivista - Articolo scientifico
catheter ablation; fluoroscopy; radiation exposure; tachycardia, atrioventricular nodal reentry; tachycardia, supraventricular; workflow; X-rays
English
2018
11
3
e005592
reserved
Pani, A., Belotti, G., Bonanno, C., Bongiorni, M., Bottoni, N., Brambilla, R., et al. (2018). Predictors of Zero X-Ray Ablation for Supraventricular Tachycardias in a Nationwide Multicenter Experience. CIRCULATION. ARRHYTHMIA AND ELECTROPHYSIOLOGY, 11(3) [10.1161/CIRCEP.117.005592].
File in questo prodotto:
File Dimensione Formato  
CIRCEP.117.005592-1.pdf

Solo gestori archivio

Tipologia di allegato: Publisher’s Version (Version of Record, VoR)
Dimensione 302.65 kB
Formato Adobe PDF
302.65 kB Adobe PDF   Visualizza/Apri   Richiedi una copia

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/10281/195679
Citazioni
  • Scopus 35
  • ???jsp.display-item.citation.isi??? 32
Social impact