The AM-5600 is a new device that simultaneously monitors electrocardiogram (ECG) and noninvasive blood pressure (BP) over a 24-h period. BP readings (Korotkoff sounds and cuff air pressure) are stored into the recorder, allowing the removal of BP artifacts after a visual check. In 12 subjects with essential hypertension, we compared BP values simultaneously provided by the AM-5600 and intra-arterial recordings. At rest, noninvasive systolic BP (SBP) values were lower (5.4 +/- 4.9 mmHg) and diastolic BP (DBP) values were higher (7.3 +/- 7.3 mmHg) than were intra-arterial values. In ambulatory conditions (9 subjects), between-method discrepancies were +0.8 +/- 6.1 and +12.2 +/- 7.4 mmHg for 24-h SBP and DBP, respectively. AM-5600 underestimated 24-h intra-arterial SBP and DBP SD, but it accurately tracked intra-arterial SBP and DBP changes. Editing removed 22.1% of total readings, slightly reducing between-method discrepancies. Thus the AM-5600 provides an accurate average estimate of resting and ambulatory SBP and, for DBP, a less accurate estimate that is slightly improved by editing. The AM-5600 allows accurate description of SBP and DBP profiles and thus may be suitable to describe the abrupt BP changes accompanying a number of clinical events

Omboni, S., Parati, G., Groppelli, A., Ulian, L., Mancia, G. (1997). Performance of the AM-5600 blood pressure monitor: Comparison with ambulatory intra-arterial pressure. JOURNAL OF APPLIED PHYSIOLOGY, 82(2), 698-703 [10.1152/jappl.1997.82.2.698].

Performance of the AM-5600 blood pressure monitor: Comparison with ambulatory intra-arterial pressure

Parati, G;Groppelli, A;Mancia, G
1997

Abstract

The AM-5600 is a new device that simultaneously monitors electrocardiogram (ECG) and noninvasive blood pressure (BP) over a 24-h period. BP readings (Korotkoff sounds and cuff air pressure) are stored into the recorder, allowing the removal of BP artifacts after a visual check. In 12 subjects with essential hypertension, we compared BP values simultaneously provided by the AM-5600 and intra-arterial recordings. At rest, noninvasive systolic BP (SBP) values were lower (5.4 +/- 4.9 mmHg) and diastolic BP (DBP) values were higher (7.3 +/- 7.3 mmHg) than were intra-arterial values. In ambulatory conditions (9 subjects), between-method discrepancies were +0.8 +/- 6.1 and +12.2 +/- 7.4 mmHg for 24-h SBP and DBP, respectively. AM-5600 underestimated 24-h intra-arterial SBP and DBP SD, but it accurately tracked intra-arterial SBP and DBP changes. Editing removed 22.1% of total readings, slightly reducing between-method discrepancies. Thus the AM-5600 provides an accurate average estimate of resting and ambulatory SBP and, for DBP, a less accurate estimate that is slightly improved by editing. The AM-5600 allows accurate description of SBP and DBP profiles and thus may be suitable to describe the abrupt BP changes accompanying a number of clinical events
Articolo in rivista - Articolo scientifico
Adult; Blood Pressure; Female; Humans; Male; Middle Aged; Blood Pressure Monitoring, Ambulatory
English
1997
82
2
698
703
none
Omboni, S., Parati, G., Groppelli, A., Ulian, L., Mancia, G. (1997). Performance of the AM-5600 blood pressure monitor: Comparison with ambulatory intra-arterial pressure. JOURNAL OF APPLIED PHYSIOLOGY, 82(2), 698-703 [10.1152/jappl.1997.82.2.698].
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/10281/191502
Citazioni
  • Scopus 6
  • ???jsp.display-item.citation.isi??? 6
Social impact