BACKGROUND Twenty-four-hour finger arterial pressure (FAP) recordings show a negative bias against intrabrachial artery pressure (BAP) and the bias is greater during the night, thereby overestimating the nocturnal blood pressure dip. We have available a methodology with which to reconstruct BAP from FAP by waveform filtering (transfer function) and generalized level (bias) correction that reduces the bias for short-term blood pressure records. OBJECTIVE To investigate if this methodology also decreases the extra bias during the night, thereby yielding a better estimate of the nocturnal dip. METHODS Twenty-four-hour FAP and BAP blood pressure recordings were simultaneously obtained in eight healthy normotensive volunteers and 14 patients with hypertension (ages 19-60 years), during standardized scheduled activities. The data were analysed off-line, applying the brachial reconstruction technique (reBAP) consisting of a waveform filter and level correction. Simultaneous beats yielded systolic, diastolic and mean pressures that were averaged per 30 min, per day, per night, per activity, over the 24-h period, and for volunteers and patients separately. RESULTS Over the full 24 h, FAP systolic, diastolic and mean values for the total group differed from BAP by +1 +/- 10, -8 +/- 7 and -10 +/- 8 mmHg (mean +/- SD), respectively. Similarly, reBAPs differed by +1 +/- 11, -2 +/- 7 and -2 +/- 7 mmHg. BAPs dipped by 20 +/- 8, 13 +/- 6 and 15 +/- 6 mmHg, respectively, during the night. These dips were overestimated by +8, +4 and +4 mmHg by FAP, but not by reBAP: -1, +1 and +1 mmHg. The volunteer and the patient groups showed slight differences in results, but these were not statistically significant. CONCLUSIONS The generalized reconstruction technique to obtain near-brachial pressure from non-invasive FAP almost completely removed bias over the full 24-h day-night period and improved tracking of diurnal changes for all three blood pressure values

Westerhof, B., Guelen, I., Parati, G., Groppelli, A., Van Montfrans, G., Wieling, W., et al. (2002). Variable day/night bias in 24-h non-invasive finger pressure against intrabrachial artery pressure is removed by waveform filtering and level correction. JOURNAL OF HYPERTENSION, 20(10), 1981-1986 [10.1097/00004872-200210000-00017].

Variable day/night bias in 24-h non-invasive finger pressure against intrabrachial artery pressure is removed by waveform filtering and level correction

PARATI, GIANFRANCO;
2002

Abstract

BACKGROUND Twenty-four-hour finger arterial pressure (FAP) recordings show a negative bias against intrabrachial artery pressure (BAP) and the bias is greater during the night, thereby overestimating the nocturnal blood pressure dip. We have available a methodology with which to reconstruct BAP from FAP by waveform filtering (transfer function) and generalized level (bias) correction that reduces the bias for short-term blood pressure records. OBJECTIVE To investigate if this methodology also decreases the extra bias during the night, thereby yielding a better estimate of the nocturnal dip. METHODS Twenty-four-hour FAP and BAP blood pressure recordings were simultaneously obtained in eight healthy normotensive volunteers and 14 patients with hypertension (ages 19-60 years), during standardized scheduled activities. The data were analysed off-line, applying the brachial reconstruction technique (reBAP) consisting of a waveform filter and level correction. Simultaneous beats yielded systolic, diastolic and mean pressures that were averaged per 30 min, per day, per night, per activity, over the 24-h period, and for volunteers and patients separately. RESULTS Over the full 24 h, FAP systolic, diastolic and mean values for the total group differed from BAP by +1 +/- 10, -8 +/- 7 and -10 +/- 8 mmHg (mean +/- SD), respectively. Similarly, reBAPs differed by +1 +/- 11, -2 +/- 7 and -2 +/- 7 mmHg. BAPs dipped by 20 +/- 8, 13 +/- 6 and 15 +/- 6 mmHg, respectively, during the night. These dips were overestimated by +8, +4 and +4 mmHg by FAP, but not by reBAP: -1, +1 and +1 mmHg. The volunteer and the patient groups showed slight differences in results, but these were not statistically significant. CONCLUSIONS The generalized reconstruction technique to obtain near-brachial pressure from non-invasive FAP almost completely removed bias over the full 24-h day-night period and improved tracking of diurnal changes for all three blood pressure values
Articolo in rivista - Articolo scientifico
Adult; Bias (Epidemiology); Blood Pressure; Brachial Artery; Circadian Rhythm; Diastole; Fingers; Humans; Hypertension; Middle Aged; Systole
English
2002
20
10
1981
1986
none
Westerhof, B., Guelen, I., Parati, G., Groppelli, A., Van Montfrans, G., Wieling, W., et al. (2002). Variable day/night bias in 24-h non-invasive finger pressure against intrabrachial artery pressure is removed by waveform filtering and level correction. JOURNAL OF HYPERTENSION, 20(10), 1981-1986 [10.1097/00004872-200210000-00017].
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/10281/173084
Citazioni
  • Scopus 29
  • ???jsp.display-item.citation.isi??? 30
Social impact