PURPOSE: Bicaval dual-lumen catheters allow for single-site cannulation venovenous extracorporeal gas exchange and facilitate early mobilization of patients. Using these catheters blood is drained from the superior and inferior venae cavae, pumped through a respiratory membrane, and returned into the right atrium. The insertion of these catheters is challenging as their correct positioning is fundamental to reduce recirculation and avoid severe complications. We describe here a new technique for the positioning of bicaval dual-lumen catheters. MATERIALS AND METHODS: The right internal jugular vein was percutaneously cannulated in nine sheep. The distance between skin and tricuspid valve was measured from the point of pressure change in the waveform of a Swan-Ganz catheter being retracted from the right ventricle into the right atrium. The atrium-tricuspid valve-ventricle axis was determined by observing the fluctuations of the tip of the Swan-Ganz entering the ventricle during fluoroscopy. A bicaval dual-lumen catheter was placed on the basis of these evaluations and connected to an extracorporeal respiratory support system. RESULTS: The position of the catheter was verified at necropsy approximately 18 h after insertion. In all cases the catheter was correctly placed, with the central port situated in front of the tricuspid valve. CONCLUSIONS: The described technique may help to position bicaval dual-lumen catheters for venovenous extracorporeal gas exchange without the use of transesophageal echocardiography or contrast media during fluoroscopy

Langer, T., Vecchi, V., Belenkiy, S., Cancio, L., Gattinoni, L., Batchinsky, A. (2013). Pressure-guided positioning of bicaval dual-lumen catheters for venovenous extracorporeal gas exchange. INTENSIVE CARE MEDICINE, 39(1), 151-154 [10.1007/s00134-012-2751-5].

Pressure-guided positioning of bicaval dual-lumen catheters for venovenous extracorporeal gas exchange

Langer, T
;
2013

Abstract

PURPOSE: Bicaval dual-lumen catheters allow for single-site cannulation venovenous extracorporeal gas exchange and facilitate early mobilization of patients. Using these catheters blood is drained from the superior and inferior venae cavae, pumped through a respiratory membrane, and returned into the right atrium. The insertion of these catheters is challenging as their correct positioning is fundamental to reduce recirculation and avoid severe complications. We describe here a new technique for the positioning of bicaval dual-lumen catheters. MATERIALS AND METHODS: The right internal jugular vein was percutaneously cannulated in nine sheep. The distance between skin and tricuspid valve was measured from the point of pressure change in the waveform of a Swan-Ganz catheter being retracted from the right ventricle into the right atrium. The atrium-tricuspid valve-ventricle axis was determined by observing the fluctuations of the tip of the Swan-Ganz entering the ventricle during fluoroscopy. A bicaval dual-lumen catheter was placed on the basis of these evaluations and connected to an extracorporeal respiratory support system. RESULTS: The position of the catheter was verified at necropsy approximately 18 h after insertion. In all cases the catheter was correctly placed, with the central port situated in front of the tricuspid valve. CONCLUSIONS: The described technique may help to position bicaval dual-lumen catheters for venovenous extracorporeal gas exchange without the use of transesophageal echocardiography or contrast media during fluoroscopy
Articolo in rivista - Articolo scientifico
ECMO; Acute respiratory distress syndrome; Experimental model
English
2013
39
1
151
154
reserved
Langer, T., Vecchi, V., Belenkiy, S., Cancio, L., Gattinoni, L., Batchinsky, A. (2013). Pressure-guided positioning of bicaval dual-lumen catheters for venovenous extracorporeal gas exchange. INTENSIVE CARE MEDICINE, 39(1), 151-154 [10.1007/s00134-012-2751-5].
File in questo prodotto:
File Dimensione Formato  
art%3A10.1007%2Fs00134-012-2751-5.pdf

Solo gestori archivio

Dimensione 233.68 kB
Formato Adobe PDF
233.68 kB Adobe PDF   Visualizza/Apri   Richiedi una copia

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/10281/254649
Citazioni
  • Scopus 14
  • ???jsp.display-item.citation.isi??? 15
Social impact