Background: Electrical impedance tomography (EIT) is a bedside monitoring technique of the respiratory system that measures impedance changes within the thorax. The close correlation between variations in impedance (ΔZ) and lung volumes (Vt) is known. Unless Vt is measured by an external reference (e.g., spirometry), its absolute value (in milliliters) cannot be determined; however, measurement of Vt would be useful in nonintubated subjects. Objective: To validate a simplified and feasible calibration method of EIT, which allows estimation of Vt in nonintubated subjects. Materials and Methods: We performed a prospective study on 13 healthy volunteers. Subjects breathed 10 times in a nonexpandable "calibration balloon" with a known volume while wearing the EIT belt. The relationship between ΔZ and the balloon volume was calculated (ΔZ/Vt). Subsequently, subjects were connected to a mechanical ventilator by a mouthpiece under different settings. Vt was calculated from EIT measurements (VtEIT) by means of the ΔZ/Vt coefficient and compared with the value obtained from the ventilator (Vtflow). Results: There was a close correlation between Vtflow and VtEIT (r2 = 0.89). The fit equation was VtEIT = 0.9 × Vtflow +10.1. The highest correlation was found at positive endexpiratory pressure (PEEP) 0 (mean: VtEIT = 0.93 × Vtflow) versus PEEP 8 (mean: VtEIT = 0.8 × Vtflow), p = 0.01. No differences in the fit equation were found between pressure support ventilation (PSV) 0 and PSV 8, p = 0.50. Further analysis showed no statistically significant differences between sex, height, and BMI. Conclusion: A simple and fast EIT calibration technique enables reliable, noninvasive monitoring of Vt in nonintubated subjects.

Sosio, S., Bellani, G., Villa, S., Lupieri, E., Mauri, T., Foti, G. (2019). A Calibration Technique for the Estimation of Lung Volumes in Nonintubated Subjects by Electrical Impedance Tomography. RESPIRATION, 98(3), 189-197 [10.1159/000499159].

A Calibration Technique for the Estimation of Lung Volumes in Nonintubated Subjects by Electrical Impedance Tomography

Sosio, S;Bellani, G
;
Villa, S;LUPIERI, ERMES;Foti, G
2019

Abstract

Background: Electrical impedance tomography (EIT) is a bedside monitoring technique of the respiratory system that measures impedance changes within the thorax. The close correlation between variations in impedance (ΔZ) and lung volumes (Vt) is known. Unless Vt is measured by an external reference (e.g., spirometry), its absolute value (in milliliters) cannot be determined; however, measurement of Vt would be useful in nonintubated subjects. Objective: To validate a simplified and feasible calibration method of EIT, which allows estimation of Vt in nonintubated subjects. Materials and Methods: We performed a prospective study on 13 healthy volunteers. Subjects breathed 10 times in a nonexpandable "calibration balloon" with a known volume while wearing the EIT belt. The relationship between ΔZ and the balloon volume was calculated (ΔZ/Vt). Subsequently, subjects were connected to a mechanical ventilator by a mouthpiece under different settings. Vt was calculated from EIT measurements (VtEIT) by means of the ΔZ/Vt coefficient and compared with the value obtained from the ventilator (Vtflow). Results: There was a close correlation between Vtflow and VtEIT (r2 = 0.89). The fit equation was VtEIT = 0.9 × Vtflow +10.1. The highest correlation was found at positive endexpiratory pressure (PEEP) 0 (mean: VtEIT = 0.93 × Vtflow) versus PEEP 8 (mean: VtEIT = 0.8 × Vtflow), p = 0.01. No differences in the fit equation were found between pressure support ventilation (PSV) 0 and PSV 8, p = 0.50. Further analysis showed no statistically significant differences between sex, height, and BMI. Conclusion: A simple and fast EIT calibration technique enables reliable, noninvasive monitoring of Vt in nonintubated subjects.
Articolo in rivista - Articolo scientifico
Electrical impedance tomography; Respiratory monitoring; Spontaneous ventilation; Tidal volume;
Electrical impedance tomography; Respiratory monitoring; Spontaneous ventilation; Tidal volume
English
2019
98
3
189
197
none
Sosio, S., Bellani, G., Villa, S., Lupieri, E., Mauri, T., Foti, G. (2019). A Calibration Technique for the Estimation of Lung Volumes in Nonintubated Subjects by Electrical Impedance Tomography. RESPIRATION, 98(3), 189-197 [10.1159/000499159].
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/10281/234838
Citazioni
  • Scopus 10
  • ???jsp.display-item.citation.isi??? 9
Social impact