Purpose: This study investigated the clinical application of a magnetic-resonance (MR)-guided breast biopsy (MRBB) system consisting of a nonmagnetic coaxial needle and a ferromagnetic core biopsy needle. Materials and methods: MRBB was performed on 70 breast lesions. The biopsy device consisted of a nonmagnetic 14- to 16-gauge coaxial needle and a ferromagnetic 16- to 18-gauge biopsy needle. Results: Of the 70 lesions, 29 were malignant and 41 nonmalignant. All 29 malignant lesions underwent surgery and were confirmed as malignant at final histology. Of the 41 nonmalignant lesions, 35 underwent follow-up breast MR imaging (mean, 26±19 months), which demonstrated no lesions changes; six lesions underwent surgery because of poor radiological-pathological correlation; of these 6 lesions, 3 were nonmalignant, one was borderline (lobular carcinoma in situ) and two were malignant (well-differentiated tubular carcinoma and infiltrating ductal carcinoma). Sensitivity, specificity, positive and negative predictive values and diagnostic accuracy were, respectively, 93.5%, 100%, 100%, 95.1% and 97.1% if the lobular carcinoma in situ was considered a nonmalignant histological result, and 90.6%, 100%, 100%, 92.7% and 95.7% if the lobular carcinoma in situ was considered malignant. Conclusions: MRBB with a ferromagnetic-nonmagnetic coaxial system represented an easy way to perform a biopsy procedure and was easily applicable in the routine clinical setting. © 2013 Springer-Verlag

Belloni, E., Panizza, P., Ravelli, S., De Cobelli, F., Gusmini, S., Losio, C., et al. (2013). Biopsia mammaria stereotassica RM-guidata mediante sistema coassiale misto ferromagnetico-amagnetico con aghi da 12-18 Gauge: esperienza clinica e risultati a lungo termine. LA RADIOLOGIA MEDICA, 118(7), 1137-1148 [10.1007/s11547-013-0954-7].

Biopsia mammaria stereotassica RM-guidata mediante sistema coassiale misto ferromagnetico-amagnetico con aghi da 12-18 Gauge: esperienza clinica e risultati a lungo termine

PERSEGHIN, GIANLUCA
Penultimo
;
2013

Abstract

Purpose: This study investigated the clinical application of a magnetic-resonance (MR)-guided breast biopsy (MRBB) system consisting of a nonmagnetic coaxial needle and a ferromagnetic core biopsy needle. Materials and methods: MRBB was performed on 70 breast lesions. The biopsy device consisted of a nonmagnetic 14- to 16-gauge coaxial needle and a ferromagnetic 16- to 18-gauge biopsy needle. Results: Of the 70 lesions, 29 were malignant and 41 nonmalignant. All 29 malignant lesions underwent surgery and were confirmed as malignant at final histology. Of the 41 nonmalignant lesions, 35 underwent follow-up breast MR imaging (mean, 26±19 months), which demonstrated no lesions changes; six lesions underwent surgery because of poor radiological-pathological correlation; of these 6 lesions, 3 were nonmalignant, one was borderline (lobular carcinoma in situ) and two were malignant (well-differentiated tubular carcinoma and infiltrating ductal carcinoma). Sensitivity, specificity, positive and negative predictive values and diagnostic accuracy were, respectively, 93.5%, 100%, 100%, 95.1% and 97.1% if the lobular carcinoma in situ was considered a nonmalignant histological result, and 90.6%, 100%, 100%, 92.7% and 95.7% if the lobular carcinoma in situ was considered malignant. Conclusions: MRBB with a ferromagnetic-nonmagnetic coaxial system represented an easy way to perform a biopsy procedure and was easily applicable in the routine clinical setting. © 2013 Springer-Verlag
Articolo in rivista - Articolo scientifico
Breast biopsy; Breast lesion; Breast magnetic resonance; Coaxial needle; Follow-up; Radiology, Nuclear Medicine and Imaging; Medicine (all)
Italian
2013
118
7
1137
1148
reserved
Belloni, E., Panizza, P., Ravelli, S., De Cobelli, F., Gusmini, S., Losio, C., et al. (2013). Biopsia mammaria stereotassica RM-guidata mediante sistema coassiale misto ferromagnetico-amagnetico con aghi da 12-18 Gauge: esperienza clinica e risultati a lungo termine. LA RADIOLOGIA MEDICA, 118(7), 1137-1148 [10.1007/s11547-013-0954-7].
File in questo prodotto:
File Dimensione Formato  
Radiol Med 2013.pdf

Solo gestori archivio

Dimensione 305.06 kB
Formato Adobe PDF
305.06 kB Adobe PDF   Visualizza/Apri   Richiedi una copia

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/10281/165212
Citazioni
  • Scopus 2
  • ???jsp.display-item.citation.isi??? 2
Social impact